Related Experiment Video
Updated: Sep 1, 2025

Reduced Procedure Time and Variability with Active Esophageal Cooling During Radiofrequency Ablation for Atrial Fibrillation
Published on: August 25, 2022
Esophageal safety in CLOSE-guided 50 W high-power-short-duration pulmonary vein isolation: The PREHEAT-PVI-registry
Alexander Francke1, Gregor Naumann1, Marie-Christin Weidauer1
1Helios Klinikum Pirna, Pirna, Germany.
Introduction:
Pulmonary vein isolation (PVI) using high-power-short-duration (HPSD) radiofrequency ablation (RF) is emerging as the standard of care for treatment of atrial fibrillation (AF). While procedural short-term to midterm efficacy and efficiency are very promising, this registry aims to investigate esopahgeal safety using an optimized ablation approach.
Methods:
In a single-center experience, 388 consecutive standardized first-time AF ablation were performed using a CLOSE-guided-fixed-50 W-circumferential PVI and substrate modification without intraprocedural esophageal temperature measurement. Three hundred patients underwent postprocedural esophageal endoscopy to diagnose and grade endoscopically detected esophageal lesions (EDEL) and were included in the analysis.
Results:
EDEL were detected in 35 of 300 patients (11.6%), 25 of 35 were low-grade Kansas-city-classification (KCC) 1 lesions with fast healing tendencies. Six patients suffered KCC 2a lesions, 4 patients had KCC 2b lesions (1.3% of all patients). No esophageal perforation or fistula formation was observed. Patient baseline characteristics, especially patients age, gender, and body mass index did not influence EDEL incidence. Additional posterior box isolation did not increase the incidence of EDEL. In patients diagnosed with EDEL, mean catheter contact force during posterior wall ablation was higher (11.9 ± 1.8 vs. 14.7 ± 3 g, p < .001), mean RF duration was shorter (11.9 ± 1 vs. 10.7 ± 1.2 s, p < .001), while achieved ablation index was not different between groups (434 ± 4.9 vs. 433 ± 9.5, n.s.).
Conclusion:
Incidence of EDEL after CLOSE-guided-50 W-HPSD PVI is lower compared to historical cohorts using standard-power RF settings. Catheter contact force during posterior HPSD ablation should not exceed 15 g.
More Related Videos
06:25Cooling or Warming the Esophagus to Reduce Esophageal Injury During Left Atrial Ablation in the Treatment of Atrial Fibrillation
Published on: March 15, 2020
05:43Author Spotlight: Improving Lesion Contiguity in Pulmonary Vein Isolation via Proactive Esophageal Cooling
Published on: April 19, 2024
Related Concept Videos
Esophageal Varices-I: Introduction
Esophageal Perforation-I: Introduction
The location of esophageal perforation can vary, occurring anywhere along the esophagus....
Pulmonary Embolism I: Introduction
Esophageal Perforation-II: Clinical Manifestations and Management
Clinical Manifestations:
Endoscopic Studies I: Bronchoscopy and Thoracoscopy
Bronchoscopy
Description
Bronchoscopy is a procedure that involves direct visualization of the larynx, trachea, and bronchi for diagnostic and therapeutic purposes. A flexible fiber optic or rigid bronchoscope is used to carry out the procedure. The fiber-optic bronchoscope is more frequently used due...
Endoscopic Procedures I: Esophagogastroduodenoscopy
During an EGD, the endoscope can be used to: