Related Experiment Video
Updated: Aug 31, 2025

Dual Raster-Scanning Photoacoustic Small-Animal Imager for Vascular Visualization
Published on: July 15, 2020
20-MHz phased array ultrasound transducer for in vivo ultrasound imaging of small animals
1Department of Applied Physics, The Hong Kong Polytechnic University, Hong Kong, China; Hospital Authority, Hong Kong, China.
Abstract:
In vivo ultrasound imaging with phased array transducers is of great importance for both clinical application and biomedical research. In this work, relaxor ferroelectric PMN-0.28PT single crystal with very high piezoelectric constant d33 ≥ 2000 pC/N and electromechanical coupling coefficient k33 ∼ 0.92 is used to fabricate high-frequency phased array transducers. A 128-element 20-MHz phased array transducer is successfully fabricated, and the optimized performance of -6 dB average bandwidth of ∼ 84 % and insertion loss of -43 dB are achieved. The axial and lateral imaging resolutions of the transducer are determined to be 81 µm and 243 µm, respectively. With Verasonics image platform, in vivo fisheye images are acquired, demonstrating the potential application of our developed high-frequency phased array transducer for biomedical research on small animals.
More Related Videos
Related Concept Videos
Ultrasonography
During an ultrasonography procedure, a handheld device called...
Ultrasound II: Endoscopic Ultrasound and FibroScan
Endoscopic Ultrasound (EUS):
Imaging Studies II: Ultrasonography