Clinical and polysomnographic predictors of suboptimal auto-adjusting CPAP titration in adult OSA patients: a
Ricardo L M Duarte1,2, Flavio J Magalhães-da-Silveira3, David Gozal4
1SleepLab, Laboratório de Estudo dos Distúrbios do Sono, Centro Médico BarraShopping, Avenida das Américas 4666, sala 309, Barra da Tijuca, Rio de Janeiro, 22649-900, Brazil. rlmduarte@gmail.com.
Purpose:
To examine potential clinical, demographic, anthropometric, and polysomnographic predictors of successful auto-adjusting continuous positive airway pressure (CPAP) titration for treatment of obstructive sleep apnea (OSA).
Methods:
This cross-sectional study was conducted in adults diagnosed with moderate-to-severe OSA (baseline apnea-hypopnea index [AHI] ≥ 15.0/h), who underwent auto-adjusting CPAP titration (S9 or S10 AutoSet ResMed®) in a sleep laboratory setting while wearing a nasal or pillow mask. Participants were then grouped into two groups: optimal CPAP titration (residual AHI < 5.0/h) or suboptimal CPAP titration (residual AHI ≥ 5.0/h). Multivariate logistic regression analysis was used to assess possible independent predictive factors for suboptimal CPAP titration.
Results:
A total of 1222 adults consisting of 874 subjects with optimal CPAP titration (71.5%) and 348 subjects with suboptimal CPAP titration (28.5%) were evaluated. Multivariate analysis resulted in a model with an adequate calibration (Hosmer-Lemeshow chi-square-test: 7.088; p = 0.527), with male sex, higher values of baseline AHI, therapeutic pressure (95th percentile), and mask leak (95th percentile) emerging as significant and independent predictors for suboptimal CPAP titration: adjusted odds ratio (OR): 1.456 (95% confidence interval [CI] 1.076-1.971; p = 0.015), OR: 1.009 (95% CI 1.002-1.016; p = 0.013), OR: 1.281 (95% CI 1.206-1.361; p < 0.001), and 1.035 (1.026-1.043; p < 0.001), respectively.
Conclusions:
In a large cohort of adults undergoing auto-adjusting CPAP titration due to moderate-to-severe OSA, male sex, increased values of baseline AHI, pressure requirements, and mask leak were significant predictors for less than optimal CPAP titration.
Related Concept Videos
Sleep Apnea
The condition is more prevalent among...
Heart Failure VI: Adjunct Therapies
Respiratory Assessment: Purpose and Indications
Objectives and Importance:
The primary goal of respiratory assessment is to evaluate patients at early risk of clinical deterioration. Since respiratory distress often precedes other signs of declining health, breathing patterns and sounds become a...
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation...
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is...
Acute Respiratory Failure-IV