Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
Published on: March 11, 2021
EXPERIMENTAL IN VITRO DOSIMETRY OF 223RA AND 177LU
Lukáš Ondrák1, Marek Sommer2,3, Václav Štěpán2
1Department of Nuclear Chemistry, Faculty of Nuclear Sciences and Physical Engineering, Czech Technical University in Prague, Břehová 7, 115 19 Prague 1, Czech Republic.
Abstract:
Targeted alpha therapy with radionuclides undergoing multiple alpha-particle decays is a promising method of nuclear medicine. To study the effectiveness of alpha versus beta emitters, survival of DU145 prostate cancer cells exposed to 223Ra or 177Lu was assessed. Per decay, the cells were much more sensitive to the alpha than beta emitter. However, per unit dose the sensitivities would be comparable, contrary to the well-known evidence, if the decay energy were deposited within the sample completely and homogeneously. Measurements by Timepix detectors showed about three times higher counts of alpha particles above than below the sample. After the first alpha decay of 223Ra to 219Rn, this gas likely moves upwards and its subsequent three alpha decays occur in the upper part of the sample. Correct estimation of absorbed dose is a critical issue when analysing in vitro data and when translating their results to clinical applications.
More Related Videos
09:49A Whole Body Dosimetry Protocol for Peptide-Receptor Radionuclide Therapy PRRT: 2D Planar Image and Hybrid 2D+3D SPECT/CT Image Methods
Published on: April 24, 2020
09:44Pretargeted Radioimmunotherapy Based on the Inverse Electron Demand Diels-Alder Reaction
Published on: January 29, 2019