Femoral Pulse Pressure Variation Is Not Interchangeable with Radial Pulse Pressure Variation during Living Donor
Doyeon Kim1, Jin Hee Ahn2, Sangbin Han3
1Department of Anesthesiology and Pain Medicine, Inha University Hospital, Inha University School of Medicine, Incheon 22332, Korea.
Abstract:
The radial artery is commonly used as the site measuring pulse pressure variation (PPV) during surgery. Accurate measurement of circulating blood volume and timely interventions to maintain optimal circulating blood volume is important to deliver sufficient oxygen to tissues and organs. It has not rather than never studied in patients undergoing liver transplantation whether PPV measured at peripheral sites, such as the radial artery, do represent central PPV for evaluating blood volume. In this retrospective study, 51 liver transplant recipients were enrolled. The two PPVs had been automatically recorded every minute in electrical medical records. A total 1878 pairs of the two PPVs were collected. The interchangeability of PPV measured at the radial and the femoral artery was analyzed by using the Bland−Altman plot, four-quadrant plot, Cohen’s kappa (k), and receiver operating curve. The bias and limits of agreement of the two PPVs were −1.3% and −8.8% to 6.2%, respectively. The percentage error was 75%. The concordance rate was 65%. The Kappa of PPV-radial determining whether PPV-femoral was >13% or ≤13% was 0.64. We found that PPV-radial is not interchangeable with PPV-femoral during liver transplantation. Additionally, PPV-radial failed to reliably track changes of PPV-femoral. Lastly, the clinical decision regarding blood volume status (depletion or not) is significantly different between the two PPVs. Therefore, PPV-femoral may help maintain blood volume circulating to major organs including the newly transplanted liver graft for liver transplant recipients.
Related Concept Videos
Sites for measruring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Assessment of radial pulse
The radial pulse, located at the wrist, is often the preferred site for assessing peripheral pulse because of its accessibility and dependability. The process of determining the radial pulse involves several steps:
Pre-Procedural Guidelines for Assessing Blood Pressure
Special considerations while measuring pulse
Pulse
The pulse serves as a clinical...
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.