Dual study on the sum-of-segments method for axial length measurement: is it better?
1From the Copenhagen Eye Center, Copenhagen, Denmark; University Eye Clinic, Aarhus, Denmark.
Purpose:
To evaluate the sum-of-segments (SOS) method for optical axial length (AL) measurements.
Setting:
Department of Ophthalmology, University of Aarhus, Denmark, and Private practice, Copenhagen, Denmark.
Design:
Retrospective observational study.
Methods:
2 retrospective datasets were included. The first dataset comprised 1491 university cataract cases measured with the LENSTAR LS900 preoperatively and 1 to 4 months postoperatively. The second dataset comprised 904 lens surgery cases with refractive follow-up to study the accuracy of intraocular lens power calculation. The prediction accuracy was evaluated as the difference between the observed and the expected refraction.
Results:
The mean difference between the preoperative and the postoperative AL readings was -0.06 mm and -0.020 mm for the standard and the SOS AL method, respectively, however with a larger variation for the SOS AL method ( P < .01). For the second dataset, the SOS method was found to increase the accuracy of the SRK/T and the Holladay formulas. With the Olsen formula, the SOS method was found to be worse ( P < .01). The highest accuracy was found using standard AL with the Olsen formula, with a mean absolute error of 0.24 diopter (D) and 89.8% of the cases within ±0.5 D.
Conclusions:
The SOS method improved the accuracy of the classical formulas probably because the optical path is a better representative of the true AL. The Olsen formula already incorporates an optical path correction, and this may be the reason for the lack of improvement with the SOS method.
More Related Videos
12:59Three and Four-Dimensional Visualization and Analysis Approaches to Study Vertebrate Axial Elongation and Segmentation
Published on: February 28, 2021
06:26Stereo-Imaging System DLT Calibration to Capture 3D In Situ Displacements of Stretched Peripheral Nerves
Published on: January 12, 2024
Related Concept Videos
Design Example: Measuring Distance Between Two Points with Obstructions
Adjusting a Traverse
Eccentric Axial Loading in a Plane of Symmetry
Beams with Unsymmetric Loadings
The first moment-area theorem determines the slope at any point on the beam. This theorem indicates that the change in slope between two points on a beam...
Beams with Symmetric Loadings
The M/EI...