Comparing a knowledge-based 3D reconstruction algorithm to TomTec 3D echocardiogram algorithm in measuring left
Attila Ahmad1,2, Sachie Shigemitsu1, Yozo Termachi1
1Division of Cardiology, Department of Pediatrics, University of Alberta, Edmonton, Alberta, Canada.
Echocardiography (Mount Kisco, N.Y.)
|August 27, 2022
Summary
This study found that Ventripoint Medical System Plus (VMS3.0+) and Tomtec software show strong agreement for measuring cardiac volumes in children using three-dimensional echocardiography (3DE). VMS3.0+ also offers faster analysis times.
Area of Science:
- Cardiology
- Medical Imaging
- Pediatric Health
Background:
- Three-dimensional echocardiography (3DE) is advancing cardiac volume measurement.
- Vendor-neutral analysis software for 3DE is limited.
- Ventripoint Medical System Plus (VMS3.0+) uses an MRI database for ventricular and atrial morphology.
Purpose of the Study:
- To compare left ventricular (LV) and left atrial (LA) volumes measured by VMS3.0+ against Tomtec echocardiography analysis software.
- To evaluate the accuracy and efficiency of VMS3.0+ for pediatric cardiac imaging.
Main Methods:
- Prospective recruitment of 98 healthy pediatric controls (0-18 years).
- Acquisition of 3D DICOM datasets for LV and LA.
- Measurement of LV and LA volumes and ejection fractions using both VMS3.0+ and Tomtec Image Arena.
- Statistical analysis included Pearson correlation, Bland-Altman plots, and intraclass coefficients (ICC).
Main Results:
- Very good correlation found between VMS3.0+ and Tomtec for LV systolic (r²=0.88, ICC=0.89) and diastolic (r²=0.88, ICC=0.90) volumes.
- Strong correlation also observed for LA diastolic (r²=0.75, ICC=0.89) and systolic (r²=0.88, ICC=0.91) volumes.
- VMS3.0+ analysis time was significantly shorter than Tomtec for both LV and LA measurements (p < 0.001).
Conclusions:
- Knowledge-based (VMS3.0+) and 3D (Tomtec) algorithms demonstrate high agreement for 3DE-derived LA and LV volumes in pediatric patients.
- VMS3.0+ provides a faster alternative for volumetric analysis compared to Tomtec.
- These findings support the use of VMS3.0+ for accurate and efficient pediatric cardiac volume assessment.