Related Experiment Video
Updated: Aug 30, 2025

Ultrasound Cyclo Plasty in Eyes with Glaucoma
Published on: January 26, 2018
Three-Year Outcomes of a Schlemm Canal Microstent (Hydrus Microstent) with Concomitant Phacoemulsification in
Ali Salimi1, Rabea Kassem2, Sangeetha Santhakumaran3
1Department of Ophthalmology, Faculty of Medicine, McGill University, Montreal, Quebec, Canada; Montreal Glaucoma Institute and Bellevue Ophthalmology Clinics, Montreal, Quebec, Canada.
Purpose:
HORIZON trials have reported the medium- and long-term outcomes of the Hydrus microstent. However, outside of randomized clinical trials, outcomes beyond 24 months have not been previously reported. Here, we examine the 3-year outcomes of the Hydrus microstent with cataract surgery outside of a randomized clinical trial setting.
Design:
Single-surgeon consecutive case series.
Participants:
Glaucomatous eyes that underwent implantation of a Hydrus microstent with concomitant cataract surgery at a single ophthalmology center in Montreal, Canada.
Methods:
Success was defined according to the absence of specific failure criteria: (A) glaucoma reoperation; (B) selective laser trabeculoplasty; (C) intraocular pressure (IOP) < 5 mmHg, > 18 mmHg, or increase in the number of antiglaucoma medications (AGMs) used (after the first postoperative month), or loss of light perception due to glaucoma; (D) aggregation of criteria A-C. Predictors of treatment failure and postoperative changes in IOP and AGM use were assessed. Safety included best-corrected visual acuity (BCVA), cup-to-disc ratio (CDR), retinal nerve fiber layer (RNFL) thickness, ganglion cell inner plexiform layer (GC-IPL) thickness, visual field mean deviation (VF-MD) and pattern standard deviation (VF-PSD), and adverse events.
Main Outcome Measures:
Surgical success, predictors of treatment failure, postoperative changes in IOP and AGM use, and structural and functional markers of disease stability.
Results:
One hundred six eyes were included. The 3-year surgical success according to Criteria A to D was 86%, 83%, 91%, and 67%, respectively. Intraocular pressure decreased by 26.5% from 18.9 ± 4.8 to 13.9 ± 2.3 mmHg (P < 0.001) and AGM use reduced by 33% from 3.0 ± 1.2 to 2.0 ± 1.2 medications (P < 0.001). The postoperative improvement in BCVA was preserved (P < 0.001). The structural and functional markers of disease stability, including CDR, RNFL, and GC-IPL thickness, as well as VF-MD and VF-PSD, remained stable (P > 0.05). Postoperative adverse events were few and transient.
Conclusions:
The 36-month results from this study show that the Hydrus microstent with phacoemulsification is safe and effective in reducing the IOP and AGM among patients with mild to severe open-angle glaucoma and can slow down the disease progression by preserving both structural and functional parameters.
Financial Disclosure(S):
Proprietary or commercial disclosure may be found after the references.
Related Concept Videos
Angle Closure Glaucoma: Treatment
Open Angle Glaucoma: Treatment
Drugs such as carbonic anhydrase inhibitors, α2- and...
Glaucoma: Overview