Performance of the imPulse device for the detection of atrial fibrillation in hospital settings
Sinéad T J McDonagh1, Shelley Rhodes2, Fiona C Warren1,2
1Primary Care Research Group, Exeter, United Kingdom.
Background:
Atrial fibrillation (AF) increases thromboembolism and stroke risk; this can be reduced by oral anticoagulation, but only if AF is detected. A portable, point-of-care device, capable of accurately detecting and identifying AF, could reduce workload and diagnostic delay by minimizing need for follow-up 12-lead electrocardiogram (ECGs).
Objective:
To assess the diagnostic performance of the Plessey imPulse lead I ECG device compared with a 12-lead ECG in detecting AF.
Methods:
Cross-sectional diagnostic accuracy study. Participants underwent simultaneous 12-lead ECG and imPulse device recordings. The imPulse device reports AF to be "probable," "possible," "unlikely," or "uncontrolled AF unlikely." imPulse and ECG reference results were cross-tabulated; sensitivity, specificity, positive/negative predictive values, and positive/negative likelihood ratios with 95% confidence interval (CI) were estimated based on different imPulse device report categorizations and heart rate subgroups.
Results:
A total of 217 participants were recruited (mean age 70.2 [standard deviation 12.7]), 56% male, 57% outpatients, 43% inpatients) and 199 were included in analyses. AF was diagnosed on ECG for 41 of 199 (20.6%) participants and reported by imPulse as possible, probable, or uncontrolled AF unlikely present for 49 of 199 (24.6%). Sensitivity and specificity for imPulse detection of possible, probable, or uncontrolled AF unlikely vs unlikely, compared with ECG, were 80.5% (95% CI, 65.1%-91.2%) and 89.9% (84.1%-94.1%), respectively. When probable or uncontrolled AF unlikely were compared vs possible or unlikely AF, sensitivity and specificity were 63.4% (46.9%-77.9%) and 98.1% (94.6%-99.6%), respectively.
Conclusion:
The imPulse device has moderate sensitivity and good specificity compared with ECG AF detection in a hospital setting.
More Related Videos
05:03Patient Directed Recording of a Bipolar Three-Lead Electrocardiogram using a Smartwatch with ECG Function
Published on: December 11, 2019
08:10Estimating Bilateral Atrial Function by Cardiovascular Magnetic Resonance Feature Tracking in Patients with Paroxysmal Atrial Fibrillation
Published on: July 20, 2022
Related Concept Videos
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Dysrhythmias V: Evaluating Dysrhythmias
Cardiopulmonary Resuscitation III: AED Use
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Disturbances in Heart Rhythm
Arrhythmias are categorized by their speed, rhythm, and origin. A slow heart...
Instrumentation Amplifier
To overcome this challenge, an ECG machine utilizes an instrumentation amplifier. This specialized amplifier is...