Evaluation of the Toa E-5000 Automated Hematology Analyzer
American Journal of Clinical Pathology
|July 1, 1987
Summary
The Toa E-5000 Automated Hematology Analyzer demonstrates excellent performance for linearity, precision, and accuracy. Erythrocyte and platelet parameters correlate well with other systems, though sample storage requires refrigeration for stability beyond 12 hours.
Area of Science:
- Clinical diagnostics
- Hematology automation
Background:
- Automated hematology analyzers are crucial for efficient and accurate blood cell counting.
- The Toa E-5000 is a novel automated hematology analyzer requiring thorough performance evaluation.
Purpose of the Study:
- To conduct a comprehensive, three-phase evaluation of the Toa E-5000 Automated Hematology Analyzer.
- To assess its analytical performance, compare key parameters with a reference analyzer, and evaluate its leukocyte differential count.
Main Methods:
- Phase 1: Assessed linearity, carryover, precision, accuracy, sample stability, and mechanical functions.
- Phase 2: Compared erythrocyte and platelet parameters against the Coulter Counter Model S-Plus IV.
- Phase 3: Evaluated leukocyte differential counts against four other methods in 300 patients (results published separately).
Main Results:
- The Toa E-5000 exhibited excellent performance characteristics in the initial evaluation.
- Sample stability was maintained at room temperature for up to 12 hours; refrigeration was needed for longer storage.
- Good correlation was observed for erythrocyte and platelet parameters with the Coulter Counter, with a noted difference in erythrocyte distribution width calculation.
Conclusions:
- The Toa E-5000 Automated Hematology Analyzer shows strong analytical performance.
- Appropriate sample handling (refrigeration after 12 hours) is recommended for maintaining parameter stability.
- The analyzer shows promise for routine hematological analysis, with further validation of differential counts available in separate publications.