Comparison between cuff-based and invasive systolic blood pressure amplification
Tan V Bui1, Dean S Picone1, Martin G Schultz1
1Menzies Institute for Medical Research, University of Tasmania, Hobart, Australia.
Cuff-based systolic blood pressure (BP) amplification measurements show bias and variability compared to invasive methods. Further research is needed to improve the accuracy of non-invasive BP monitoring devices.
Area of Science:
- Cardiovascular physiology
- Medical device technology
- Hypertension research
Background:
- Accurate central blood pressure (BP) measurement is crucial for cardiovascular risk assessment.
- Systolic blood pressure (SBP) amplification, the difference between brachial and central SBP, is an important hemodynamic parameter.
- Upper arm cuff-based methods are widely used for BP measurement, but their accuracy in determining SBP amplification needs validation against invasive techniques.
Purpose of the Study:
- To evaluate the agreement between cuff-based and invasively measured systolic blood pressure (SBP) amplification.
- To assess the performance of two different central BP devices in estimating SBP amplification.
Main Methods:
- Simultaneous invasive and cuff-based SBP amplification measurements were performed in patients undergoing coronary angiography.
- Two central BP devices were used: Sphygmocor Xcel (device 1, n=171) and Uscom BP+ (device 2, n=52).
- Statistical analysis included calculating mean differences, limits of agreement, and concordance across SBP amplification quartiles.
Main Results:
- Device 1 showed a mean difference of 4 mmHg (-12, +20 mmHg) and Device 2 showed -2 mmHg (-14, +10 mmHg) compared to invasive measurements.
- Both devices exhibited systematic bias, overestimating SBP amplification at lower levels and underestimating at higher levels.
- Concordance was low, especially in the lowest and highest quartiles of invasive SBP amplification; however, cuff-based estimates had less individual variability.
Conclusions:
- Cuff-based SBP amplification measurements demonstrate proportional systematic bias when compared to invasive methods.
- Despite biases, cuff-based methods exhibit lower individual variability, suggesting potential for improvement in non-invasive central BP monitoring.
- Findings provide insights for enhancing the performance of cuff-based central BP measurement devices.
More Related Videos
Related Concept Videos
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Sites for measruring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Assessment of blood pressure in brachial artery(two-step method)
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Measurement of Blood Pressure
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment: