Related Experiment Video
Updated: Aug 29, 2025

Patient Directed Recording of a Bipolar Three-Lead Electrocardiogram using a Smartwatch with ECG Function
Published on: December 11, 2019
Accuracy of mobile 6-lead electrocardiogram device for assessment of QT interval: a prospective validation study
A T Bergeman1,2, S N J Pultoo1, M M Winter1,2
1Department of Clinical and Experimental Cardiology, Heart Centre, Amsterdam Cardiovascular Sciences, Amsterdam University Medical Centres, location Academic Medical Centre, University of Amsterdam, Amsterdam, The Netherlands.
Introduction:
Ambulatory assessment of the heart rate-corrected QT interval (QTc) can be of diagnostic value, for example in patients on QTc-prolonging medication. Repeating sequential 12-lead electrocardiograms (ECGs) to monitor the QTc is cumbersome, but mobile ECG (mECG) devices can potentially solve this problem. As the accuracy of single-lead mECG devices is reportedly variable, a multilead mECG device may be more accurate.
Methods:
This prospective dual-centre study included outpatients visiting our cardiology clinics for any indication. Participants underwent an mECG recording using a smartphone-enabled 6‑lead mECG device immediately before or immediately after a conventional 12-lead ECG recording. Multiple QTc values in both recordings were manually measured in leads I and II using the tangent method and subsequently compared.
Results:
In total, 234 subjects were included (mean ± standard deviation (SD) age: 57 ± 17 years; 58% males), of whom 133 (57%) had cardiac disease. QTc measurement in any lead was impossible due to artefacts in 16 mECGs (7%) and no 12-lead ECGs. Mean (± SD) QTc in lead II on the mECG and 12-lead ECG was 401 ± 30 and 406 ± 31 ms, respectively. Mean (± SD) absolute difference in QTc values between both modalities was 12 ± 9 ms (r = 0.856; p < 0.001). In 55% of the subjects, the absolute difference between QTc values was < 10 ms.
Conclusion:
A 6-lead mECG allows for QTc assessment with good accuracy and can be used safely in ambulatory QTc monitoring. This may improve patient satisfaction and reduce healthcare costs.
More Related Videos
Related Concept Videos
Electrocardiogram
Three major waveforms are present in a typical ECG recording: the P wave, the QRS complex, and...
Electrocardiogram Fundamentals
An electrocardiogram (ECG) is a diagnostic tool for identifying cardiac conditions such as arrhythmias, conduction abnormalities, and myocardial ischemia.
Definition
An electrocardiogram (ECG) visualizes the heart's electrical activity by tracing the electrical movement associated with each heartbeat on a graph or monitor. As the heart beats, an electrical wave passes through it, correlating with the cardiac cycle events.
Parts of an ECG
An ECG utilizes electrodes on the skin...
ECG Interpretation of Rhythms
Components of the Electrocardiogram
The primary components of a normal ECG waveform in Normal sinus rhythm(NSR) include the P wave, PR interval, QRS complex, ST segment, T wave, and occasionally a U wave.
ECG waveforms are divided by vertical and horizontal lines at standard intervals.
The horizontal axis measures time and rate, and the vertical axis measures amplitude or voltage....
Holter Monitor: 24-Hour Monitoring
Dysrhythmias V: Evaluating Dysrhythmias
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...