Purpose:
Preprocedural computed tomography (CT) imaging before transcatheter aortic valve implantation/replacement (TAVI/TAVR) requires high diagnostic accuracy without motion artifacts. The aim of this retrospective study is to compare the image quality of a high-pitch non-electrocardiography (ECG)-gated CT protocol used in patients with atrial tachyarrhythmias with a prospectively ECG-gated CT protocol used in patients with sinus rhythm.
Materials And Methods:
We retrospectively included 108 patients who underwent preprocedural CT imaging before TAVI/TAVR. 52 patients with sinus rhythm were imaged using a prospectively ECG-gated protocol (Group A), and 56 patients with atrial tachyarrhythmias were imaged using the high-pitch non-ECG-gated protocol (Group B). Image quality was rated subjectively by two experienced radiologists and assessed by objective parameters including radiation dose, image noise, contrast-to-noise ratio (CNR), and signal-to-noise ratio (SNR) at the levels of the aortic root and abdominal aorta.
Results:
Subjective image quality was equally good with both CT protocols, and interrater agreement was substantial in both groups but tended to be higher in Group B at the level of the aortic root (Group A: κw = 0.644, Group B: κw = 0.741). With the high-pitch non-ECG-gated CT protocol, image noise was significantly increased (p = 0.001), whereas the SNR, CNR, and radiation dose were significantly decreased (p = 0.002, p = 0.003, and p < 0.001, respectively) at the level of the aortic root compared to the prospectively ECG-gated CT protocol.
Conclusion:
The high-pitch non-ECG-gated protocol yields images with similar subjective image quality compared with the prospectively ECG-gated CT protocol and allows motion-free assessment of the aortic root for accurate TAVI/TAVR planning. The high-pitch non-ECG-gated protocol may be used as an alternative for preprocedural CT imaging in patients with atrial tachyarrhythmias.
Key Points:
· In patients with atrial tachyarrhythmias, a high-pitch non-ECG-gated CT protocol achieves similar subjective image quality compared to a prospective ECG-gated CT protocol.. · At the level of the aortic root, image noise is significantly increased, whereas SNR and CNR are significantly decreased using the high-pitch non-ECG-gated protocol.. · Radiation dose is reduced by 55 % using the high-pitch non-ECG-gated protocol..
Citation Format:
· Shnayien S, Beetz N, Bressem KK et al. Comparison of a High-Pitch Non-ECG-Gated and a Prospective ECG-Gated Protocol for Preprocedural Computed Tomography Imaging Before TAVI/TAVR. Fortschr Röntgenstr 2023; 195: 139 - 147.
More Related Videos
Related Concept Videos
Imaging Studies for Cardiovascular System V: CT
Cardiac Catheterization I: Pre-Procedure Overview
Imaging Studies for Cardiovascular System II:Types of Echocardiography
Types of Echocardiography
Transthoracic Echocardiography (TTE)
TTE is the most common type of echocardiogram which involves placing a transducer on the patient's chest, emitting sound waves to create heart images. TTE is invaluable for evaluating the heart's size, structure, and motion, making it particularly useful for...
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Imaging Studies III: Computed Tomography