Repeatability and comparability of a new swept-source optical coherence tomographer in optical biometry
C E Monera Lucas1, J Escolano Serrano2, C Tarazona Jaimes1
1Servicio de Oftalmología, Hospital General Universitario de Elche, Alicante, Spain; Programa de Doctorado en Salud Pública, Ciencias Médicas y Quirúrgicas, Universidad Miguel Hernández de Elche, Alicante, Spain.
Archivos De La Sociedad Espanola De Oftalmologia
|September 6, 2022
Summary
The new Anterion swept-source OCT device provides reliable and reproducible ocular biometry measurements. Its performance is comparable to the Lenstar LS900, ensuring accuracy for cataract patients.
Area of Science:
- Ophthalmology
- Medical Devices
- Biometry
Background:
- Accurate ocular biometry is crucial for refractive outcomes in cataract surgery.
- New devices require validation for precision and comparability with established methods.
Purpose of the Study:
- To assess the reproducibility of ocular biometric measurements using the Anterion swept-source OCT (SS-OCT).
- To compare the Anterion SS-OCT's measurements with those from the Lenstar LS900 optical low coherence reflectometry biometer.
Main Methods:
- An observational study included 45 cataract patients.
- Ocular biometric parameters (axial length, anterior chamber depth, etc.) were measured thrice with Anterion and once with Lenstar LS900.
- Statistical analyses included Pearson's R, ICC, and CCC to evaluate repeatability and comparability.
Main Results:
- Excellent repeatability (Pearson's R) was observed for axial length (0.987), central corneal thickness (0.983), and anterior chamber depth (0.942).
- No significant differences were found between repeated Anterion measurements.
- Excellent agreement (ICC, CCC) was noted for axial length, central corneal thickness, and anterior chamber depth.
Conclusions:
- Biometry using the Anterion SS-OCT is a reliable and reproducible procedure.
- The Anterion SS-OCT demonstrates comparability with the Lenstar LS900 for ocular biometry.