X-ray Dose Reduction through Adaptive Exposure in Fluoroscopic Imaging
Published on: September 11, 2011
Performance of automatic exposure control on dose and image quality: comparison between slot-scanning and flat-panel
Guillaume Boissonnat1, Pierre Morichau-Beauchant2, Aymeric Reshef2
1Université Paris-Saclay, CEA, List, Palaiseau, France.
Background:
EOSedge™* (EOS Imaging, Paris, France) is an X-ray imaging system using automatic exposure control (AEC) with tube current modulation, in order to optimize dose deposition in patients.
Purpose:
This study aims at characterizing EOSedge organ dose deposition in comparison to a digital radiography (DR) system and the previous EOS system (EOS-1st generation), in relation to their respective image quality levels.
Method:
Organ doses were measured in an anthropomorphic female adult phantom and a 5-year-old pediatric phantom using optically stimulated luminescence (OSL) dosimeters, which were carefully calibrated within the studied energy range. Organ doses were recorded on the EOSedge and the Fuji Visionary DRF (Fujifilm Medical Systems U.S.A., Inc, Lexington, MA). The resulting effective doses were compared to the EOS-1st-generation values present in the literature. Image quality assessment was carried out on end-user images. Quantitative image quality metrics were computed for all tested modalities on a quality assurance phantom. Qualitative assessment of EOSedge image quality was based on anthropomorphic phantom acquisitions against the EOS-1st-generation system, and on clinical images against the tested DR system.
Results:
For a full-spine exam, and on the female adult phantom (respectively, the pediatric phantom), an effective dose of 92 μSv (respectively, 32 μSv) was obtained on EOSedge, and 572 μSv (respectively, 179 μSv) on the DR system; these values were compared to effective dose values of 290 μSv (respectively, 200 μSv) from the literature on EOS-1st generation, leading to an effective dose reduction factor of 6 with respect to the DR system, and of 3-6 with respect to EOS-1st generation. EOSedge provides the best compromise between contrast-to-noise ratio (CNR) and dose, with more consistent CNR values than the other tested modalities, in a range of attenuation from 10 to 40 cm of poly(methyl methacrylate) (PMMA). Within this range, EOSedge is also comparable to DR for 10 and 20 cm of PMMA, and better than DR for 30 and 40 cm of PMMA, both in terms of spatial resolution and low-contrast detection. The anatomical landmarks of interest in the follow-up of spinal deformities can be detected in all tested modalities.
Conclusion:
Results showed that EOSedge provides significant dose reduction factors for full spine imaging in both adults and children compared to the other tested modalities, without compromising image quality. We believe that this work could help raise awareness on the capabilities of modern X-ray systems, when equipped with appropriate AEC strategies, to perform ultra-low-dose, long-axis images.
More Related Videos
06:20Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
Published on: March 11, 2021
07:57Positron Emission Tomography-based Dose Painting Radiation Therapy in a Glioblastoma Rat Model using the Small Animal Radiation Research Platform
Published on: March 24, 2022
Related Concept Videos
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Radiological Investigation I: X-ray and CT
X-ray Imaging
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies for Cardiovascular System V: CT