Evaluation of Common Nasal Cannulas in Neonatal Noninvasive Ventilation (NIV) Using a Novel Neonatal Nasal Model
Ulf Borg1, Jeffrey Aviano2, Milan Ginani2
1Department of Medical Science, Respiratory Interventions, Medtronic, Boulder, CO, USA.
Purpose:
Non-invasive ventilation (NIV) may reduce intubation rates and be especially beneficial in the care of preterm infants, in addition to other care modalities. Currently, ventilators do not display the pressure at the nares but the pressure in the ventilator tubing system. There are several nasal cannulas available for use to deliver NIV. The purpose of this study was to compare the inspiratory pressure on the ventilator to the measured pressure delivered at the nares using three cannula brands (Manufacturer A, Fisher & Paykel; Manufacturer B, Neotech RAM; and Manufacturer C, Hudson RCI).
Patients And Methods:
This bench study utilized a 3D printed nasal model, including nares in multiple sizes to accommodate all nasal prongs studied. The nasal adaptors were connected to neonatal test lungs, to simulate patient breathing. Multiple sizes of nasal cannulas from the three manufacturers were tested for inspiratory vs delivered pressure at the patient side of the cannula, using eight combinations of ventilator settings. Each nasal cannula was tested on six Puritan Bennett™ 980 ventilators.
Results:
The measured delivered pressure at the nares was consistently lower than the clinician-set inspiratory pressure. Across all ventilator settings, 7 of the 11 cannulas delivered significantly less pressure at the nares compared to the inspiratory ventilator pressure (p < 0.01). For each cannula, as inspiratory pressure increased, the difference between delivered and inspiratory pressures also increased. The cannula from Manufacturer B consistently demonstrated the greatest differences between set inspiratory and delivered pressures for each ventilator setting.
Conclusion:
This study demonstrated substantial differences between ventilator inspiratory pressure and measured delivered pressure, which may have clinical significance. Being unaware of the actual airway pressure delivered to the patient may lead to erroneous adjustments to the level of ventilator pressure, which may be especially consequential to those with delicate and developing respiratory systems.
More Related Videos
04:56Laryngeal Mask Airway LMA Placement in a Neonatal Patient Simulator Using a Non-Inflatable Supraglottic Airway SGA
Published on: July 14, 2023
15:04Modeling and Simulations of Olfactory Drug Delivery with Passive and Active Controls of Nasally Inhaled Pharmaceutical Aerosols
Published on: May 20, 2016
Related Concept Videos
Administering Oxygen by Nasal Cannula
Nasal Cannulas
A nasal cannula is a lightweight tube split into two prongs placed in the nostrils,...
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Suctioning the Nasopharyngeal Airway
Equipment Required
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation...