Biomechanical behaviour of tension-band-reconstruction titanium plate in open-door laminoplasty: a study based on finite element analysis
Related Experiment Video
Updated: Aug 29, 2025

Author Spotlight: Development of a Novel Finite Element Analysis Model for Improved Orthognathic Surgical Techniques
Published on: October 20, 2023
Hanpeng Xu1, Jincheng Wu1, Hongru Xie1
1Graduate School, Tianjin Medical University, Tianjin, China.
Objective:
To investigate and evaluate the biomechanical behaviour of tension-band-reconstruction (TBR) and ordinary titanium plates in open-door laminoplasty by finite element (FE) analysis.
Methods:
TBR titanium plate and ordinary titanium plate were implanted into a validated finite element model of healthy adult cervical vertebrae. Among them, 5 ordinary titanium plate were used in model A, 2 TBR titanium plates and 3 ordinary titanium plates were used in model B, and 5 TBR titanium plates were used in model C. The same loading conditions was applied identically to all models. Range of motion (ROM) of the vertebral body, stress distribution of the titanium plate and intradiscal pressure (IDP) were compared in flexion, extension, lateral bending and rotation.
Results:
The ROM of model B and C was similar in flexion and extension, and both were smaller than that of model A. The highest von Mises stress in the titanium plate appears is in model C. The IDP in C2/3 was significantly higher than that in other segments in flexion. There was no significant difference in IDP among three models in left lateral bending and left axial rotation.
Conclusion:
Application of TBR titanium plate in open-door laminoplasty can reduced ROM in flexion, extension and axial rotation of the cervical vertebrae. But the increase of stress in TBR titanium plate could lead to higher risk of adverse events such as titanium plate deformation. Moreover, compared with complete TBR titanium plate, the combination of TBR titanium plate for C3 and C7 with ordinary titanium plate for the other vertebrae largely reduce the stress of the titanium plates by ensuring stability. The proposed FE model (C2-T1) exhibits a great potential in evaluating biomechanical behaviour of TBR titanium plate for open-door laminoplasty.
More Related Videos
15:11Surgical Fixation of Sternal Fractures: Preoperative Planning and a Safe Surgical Technique Using Locked Titanium Plates and Depth Limited Drilling
Published on: January 5, 2015
06:38Establishment of a Segmental Femoral Critical-size Defect Model in Mice Stabilized by Plate Osteosynthesis
Published on: October 12, 2016
Related Concept Videos
Members Made of Elastoplastic Material
As the bending moment...
Plastic Deformations of Members with a Single Plane of Symmetry