Accuracy of transcutaneous bilirubin level measured by a JM-105 bilirubinometer
Akira Ohishi1, Miyuki Kondo1, Tomoka Fujita1
1Maternal-Fetal and Neonatal Care Center, Hamamatsu University School of Medicine, Hamamatsu, Japan.
Background:
Transcutaneous bilirubin (TcB) measurement is useful, but dissociation with total serum bilirubin (TSB) is a clinical problem in measurement. We verified the accuracy of the latest version of the JM-105 jaundice meter.
Methods:
The TcB, TSB, and hematocrit (Hct) measurements obtained in the first 4 days of life in 2788 term neonates were analyzed.
Results:
When divided into 2-mg/dL classes, the difference between the TcB and TSB measurements did not change as TcB increased, but both overestimation and underestimation of TcB increased as TcB increased. At TcB greater than 11 mg/dL, inaccurate measurements with dissociation greater than 2 mg/dL exceeded 10% of the TcB measurements. The Hct value was associated with overestimation and underestimation.
Conclusion:
To evaluate neonatal jaundice accurately, it is desirable to measure TSB by blood sampling before discharge from obstetrics or in the case of worsening jaundice on day 4 or 5 of life.
Related Concept Videos
Uncertainty in Measurement: Reading Instruments
Assessing Body Temperature - Rectal
Follow these steps for rectal temperature assessment:
Step 1: Perform hand hygiene and don clean gloves to prevent cross-infection.
Step 2: Position the patient in a side-lying position to better visualize the rectal...
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Assessment of blood pressure in brachial artery(two-step method)
Measurement of Blood Pressure
Drug Concentrations: Measurements
Plasma...