Description and validation of an LDL procedure designed to select SSPL90
Ear and Hearing
|June 1, 1987
Summary
A new method measures loudness discomfort levels (LDLs) for selecting hearing aid maximum output. This procedure provides practical selection of SSPL90 using sound pressure levels (SPL) in a 2 cm3 coupler.
Area of Science:
- Audiology
- Hearing Aid Technology
- Psychoacoustics
Background:
- Selecting appropriate hearing aid maximum output (SSPL90) is crucial for patient comfort and audibility.
- Existing methods for determining SSPL90 may lack practicality or direct applicability to hearing aid fitting.
Purpose of the Study:
- To describe a novel procedure for measuring loudness discomfort levels (LDLs).
- To utilize LDL measurements for the practical selection of SSPL90 characteristics in hearing aids.
Main Methods:
- Participants were seated in a sound field wearing a high-output hearing aid with a known 2 cm3 coupler gain.
- Frequency-specific signals were presented, and participants used loudness category descriptors to rate perceived loudness.
- Loudness discomfort levels (LDLs) were defined in terms of sound pressure level (SPL) within a 2 cm3 coupler.
Main Results:
- The described procedure allows for the practical selection of SSPL90 directly from hearing aid specification sheets.
- Experiments demonstrated the stability of LDL measurements over time.
- Validation confirmed the effectiveness of the SSPL90 selection based on the new LDL measurement procedure.
Conclusions:
- The new procedure offers a practical and validated method for selecting hearing aid SSPL90.
- This approach enhances the precision and ease of hearing aid fitting by directly linking LDLs to SSPL90 selection.
- The findings support the use of coupler-based LDL measurements for optimizing hearing aid performance.