Assessment of abdominal organ dose and image quality in varying arc trajectory interventional C-arm cone beam CT
Antonios E Papadakis1, John Damilakis2
1Medical Physics Department, University General Hospital of Heraklion, Stavrakia 71110, Crete, Greece.
Purpose:
The aim of this study was to investigate the effect of varying arc exposure trajectory on radiation dose to radiosensitive organs and to assess image quality in abdominal C-arm cone beam computed tomography (CBCT) interventional procedures using a latest generation system.
Methods:
An anthropomorphic phantom that simulates the average adult individual was used. Individual-specific Monte Carlo (MC) simulation dosimetry was performed to estimate organ doses (OD) in abdominal C-arm CBCT. Seven examination protocols prescribed by the system for vascular and soft tissue CBCT, were simulated. These protocols are differentiated in the range of the arc exposure trajectory and the level of radiation dose delivered to the patient. OD was estimated for liver, adrenals, kidneys, pancreas, stomach, gall bladder, spleen, bone and skin. Image noise, signal to noise ratio (SNR), contrast to noise ratio (CNR) and in-plane spatial resolution were assessed using CT-specific image quality assessment phantoms.
Results:
OD was found to depend on the range of arc trajectory and was higher for posterior located organs. In vascular protocols OD ranged from 4.75 mGy for skin to 0.60 mGy for bone. Image noise was higher in vascular protocols than in soft tissue ones. SNR and CNR were significantly modified among different soft tissue protocols (P < 0.05). In-plane spatial resolution was found 0.80 lp/mm in vascular as opposed to 0.41 lp/mm in soft tissue protocols.
Conclusions:
The current results may be used to estimate OD for different examination protocols and enable operators choose the appropriate acquisition protocol on the preprogrammed interventional task.
More Related Videos
09:32Time-Resolved, Dynamic Computed Tomography Angiography for Characterization of Aortic Endoleaks and Treatment Guidance via 2D-3D Fusion-Imaging
Published on: December 9, 2021
05:49Author Spotlight: Advancing CBCT and Digital Dental Image Integration with AI-Assisted Digitization
Published on: February 23, 2024
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies for Cardiovascular System V: CT
Imaging Studies III: Computed Tomography
X-ray Imaging
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Radiological Investigation I: X-ray and CT