Pulsed-field ablation: Computational modeling of electric fields for lesion depth analysis
Daniel Meckes1,2, Mehrdad Emami1, Ian Fong2
1Centre for Heart Rhythm Disorders, University of Adelaide and Royal Adelaide Hospital, Adelaide, Australia.
Heart Rhythm O2
|September 13, 2022
Summary
Pulsed-field ablation (PFA) requires specific voltage and catheter contact for effective myocardial injury. Optimal catheter contact is crucial for achieving desired lesion depth, with increased distance significantly raising required voltage.
Area of Science:
- Cardiovascular research
- Medical device technology
- Electrophysiology
Background:
- Pulsed-field ablation (PFA) is a novel nonthermal cardiac ablation technology.
- The precise voltage needed for effective myocardial injury with PFA is not fully understood.
- PFA's pulsed-field strength is independent of tissue contact, leading to assumptions about contact necessity.
Purpose of the Study:
- To investigate the relationship between applied voltage, catheter-to-surface distance, and myocardial injury depth during PFA.
- To quantify the impact of these parameters on irreversible myocardial damage.
Main Methods:
- A computational model was developed and validated using in vivo data.
- The model simulated PFA to analyze varying applied voltages and catheter-tissue distances.
- The study focused on predicting the depth of irreversible myocardial injury.
Main Results:
- Lesion depth in PFA is dose-dependent, directly correlating with applied voltage.
- A minimum of 300 volts is required for lesions >1 mm deep.
- Optimal catheter contact is vital; a 1-2 mm gap increases the required voltage from 700 to 1500 volts for a 3 mm lesion.
Conclusions:
- PFA demonstrates dose-dependent effects on myocardial injury depth.
- Maintaining consistent catheter-tissue contact is critical for effective PFA outcomes.
- These findings highlight the importance of precise parameter control in PFA procedures.