Coulomb Explosion Imaging as a Tool to Distinguish Between Stereoisomers
Published on: August 18, 2017
Modified Klyshko method for an analog detector calibration.
This study extends quantum efficiency (QE) calibration to analog detectors, achieving 4% uncertainty for photomultiplier tubes (PMTs) without external references. The method accurately measures single-photon responses for improved detector characterization.
Area of Science:
- Quantum optics
- Detector physics
- Metrology
Background:
- Absolute quantum efficiency (QE) calibration is crucial for photon detectors.
- Existing methods are limited to photon counters, excluding analog detectors.
- Analog detectors exhibit a high dispersion of single-photon responses, complicating calibration.
Purpose of the Study:
- To extend absolute QE calibration to analog detectors.
- To develop a reference-free calibration method for photomultiplier tubes (PMTs).
- To achieve high accuracy in QE measurements for analog detectors.
Main Methods:
- Measuring the biphoton field correlation function.
- Approximating the distribution of detector readings.
- Determining the average number of photoelectrons per response.
- Calculating the average charge corresponding to a single photoelectron.
Main Results:
- Successful extension of absolute QE calibration to analog detectors.
- Reference-free measurement of a PMT cathode QE with 4% relative uncertainty.
- Validation through an independent experiment showing agreement with a single-photon counter.
Conclusions:
- The developed method enables accurate absolute QE calibration for analog detectors.
- This advancement is vital for applications requiring precise photon detection.
- The technique offers a reliable, reference-free approach for PMT characterization.
More Related Videos
11:08Double-barreled and Concentric Microelectrodes for Measurement of Extracellular Ion Signals in Brain Tissue
Published on: September 5, 2015
12:48Measuring Cation Transport by Na,K- and H,K-ATPase in Xenopus Oocytes by Atomic Absorption Spectrophotometry: An Alternative to Radioisotope Assays
Published on: February 19, 2013
Related Concept Videos
Instrument Calibration
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...
Difference from Background: Limit of Detection
The LOD indicates the presence or absence...
Flame Photometry: Lab
Precipitation Titration: Endpoint Detection Methods
In the Volhard method, a standard excess of AgNO3 is first added to the...
Calibration Curves: Linear Least Squares
For data that follow a straight line, the standard method for fitting is the linear...