In vitro and in vivo evaluation of a novel wired transmission magnetically controlled capsule endoscopy system for
Yaoping Zhang1,2, Lina Qu1,2, Yani Gou1,2
1Department of Gastroenterology, Lanzhou University Second Hospital, Lanzhou, Gansu Province, China.
Background:
Magnetically controlled capsule endoscopy (MCCE) has recently increasingly been used for gastric examination. However, the image quality and esophageal observation is suboptimal. We developed a novel wired transmission magnetically controlled capsule endoscopy (WT-MCCE) system and evaluated its feasibility through in vitro and in vivo experiments.
Methods:
A plastic stomach model and a pathological upper gastrointestinal model were used to evaluate the performance of WT-MCCE in vitro experiments. Twice of examination in the two in vitro models by WT-MCCE were performed by 5 endoscopists who were experienced in performing wireless capsule endoscopy. The examination of traditional gastroscopy (Olympus, GIF-HQ290) in the pathological upper gastrointestinal model was set as the control. In vivo experiments were performed in a live canine model by 3 endoscopists, in which WT-MCCE was inserted with the assistance of gastroscopy. Measurements included maneuverability, examination time, visualization of gastric mucosa, image quality and diagnostic accuracy.
Results:
WT-MCCE showed good performance in both in vitro and in vivo experiments with excellent visualization of mucosa (75-100%). The mean operation time is 17.6 ± 2.7 min, 22.3 ± 1.9 min and 29.3 ± 3.4 min in three models, respectively. In pathological upper gastrointestinal model, all lesions, including esophageal varices, one polyp, one foreign body, two gastric ulcers and one duodenal ulcer, were detected by both WT-MCCE and traditional gastroscopy by all endoscopists. For the observation of esophagus and stomach in the canine model, WT-MCCE also showed excellent maneuverability and good image quality.
Conclusions:
The novel WT-MCCE system performed well in evaluating upper gastrointestinal landmarks and lesions in two in vitro models, and showed good performance in a canine model. WT-MCCE may be potentially useful for diagnosis of esophageal and gastric diseases.
More Related Videos
09:29Fabrication and Implantation of Miniature Dual-element Strain Gages for Measuring In Vivo Gastrointestinal Contractions in Rodents.
Published on: September 18, 2014
09:42Murine Endoscopy for In Vivo Multimodal Imaging of Carcinogenesis and Assessment of Intestinal Wound Healing and Inflammation
Published on: August 26, 2014
Related Concept Videos
Endoscopic Procedures III: Video Capsule Endoscopy
Imaging Studies III: Gastrointestinal Motility Studies and Virtual Colonoscopy
Radionuclide Testing
Radionuclide testing is a sophisticated medical technique for assessing gastrointestinal motility. It focuses on gastric emptying and colonic transit time. Radioactive markers track the movement of food through the digestive system, providing insights into gastrointestinal disorders.
In gastric emptying studies, a meal's liquid and...
Urologic Endoscopic Procedure: Cystoscopic Examination
Endoscopic Procedures I: Esophagogastroduodenoscopy
During an EGD, the endoscope can be used to: