Related Experiment Video
Updated: Aug 28, 2025

Multi-Tracer Studies of Brain Oxygen and Glucose Metabolism Using a Time-of-Flight Positron Emission Tomography-Computed Tomography Scanner
Published on: June 7, 2024
PHANTOM EXPERIMENTAL STUDY ON PATIENT DOSES OBTAINED FROM 320-MULTIDETECTOR-ROW COMPUTED TOMOGRAPHY IN WHOLE-BRAIN
Heqing Lu1, Ting Meng2, Yang You2
1Department of Medical Equipment, Shanghai First Maternity and Infant Hospital, School of Medicine, Tongji University, Shanghai 200092, China.
Abstract:
This study aimed to precisely evaluate organ dose and effective dose (E) obtained from a 320-multidetector-row computed tomography (CT) scanner in brain perfusion scans and to estimate the conversion factor (k) between E and dose length product (DLP). A total of 270 thermoluminescent dosemeters were implanted in a male anthropomorphic phantom to measure air kerma. The ratios of mass-energy absorption coefficients were used to convert air kerma into organ doses. The organ doses ranged from 0.01 to 150 mGy. Doses in brain, salivary glands and red bone marrow were relatively high, and dose in eye lens reached about 110 mGy. The resulting effective dose was 5.30 mSv. The resulting conversion factor k = (0.0022 ± 0.0002) mSv·(mGy·cm)-1 was not significantly different from that of 0.0021 mSv·(mGy·cm)-1 reported for head CT scan in ICRP Publication 102.