Related Experiment Video
Updated: Aug 28, 2025

Improving IV Insulin Administration in a Community Hospital
Published on: June 11, 2012
Uncovering Discrepancies in IV Vancomycin Infusion Records between Pump Logs and EHR Documentation
Tsan-Hua Tung1, Poching DeLaurentis2, Yuehwern Yih1
1School of Industrial Engineering, College of Engineering, Purdue University, West Lafayette, Indiana, United States.
Discrepancies between electronic health record (EHR) documentation and infusion pump logs for vancomycin administration are common. These timestamp errors can impact therapeutic drug monitoring (TDM) decisions.
Area of Science:
- Pharmacokinetics and Pharmacodynamics
- Health Informatics
- Clinical Pharmacy Practice
Background:
- Accurate infusion timing is critical for vancomycin therapeutic drug monitoring (TDM), influencing area under the concentration-time curve (AUC) and trough levels.
- Current methods for assessing infusion time accuracy are resource-intensive and difficult to scale.
- Electronic health records (EHRs) may not accurately reflect actual drug administration events recorded by infusion pumps.
Purpose of the Study:
- To analyze timestamp discrepancies between EHR documentation and infusion pump event logs for vancomycin.
- To determine the prevalence of inaccurate infusion start times, completion times, and interruptions documented in EHRs.
- To highlight potential impacts on clinical decision-making for vancomycin therapy.
Main Methods:
- Utilized process mining techniques to compare EHR data with infusion pump event logs.
- Developed an algorithm to link pump event logs to corresponding EHR records for vancomycin infusions.
- Analyzed discrepancies in infusion start time, completion time, and interruption duration.
Main Results:
- Significant discrepancies were found in infusion start times (19.1% > ±10 minutes).
- A small percentage of infusion interruptions exceeded 20 minutes (2.5%).
- Completion time discrepancies were noted in a substantial portion of both 1-hour (24.2%) and 2-hour (32%) vancomycin infusions.
Conclusions:
- Infusion pump event logs and EHR data can be leveraged to study infusion administration time discrepancies at scale.
- Timestamp discrepancies are likely an inherent issue within current EHR documentation processes in hospitals.
- Further investigation across multiple institutions is recommended to understand the prevalence and guide improvement efforts.
Related Concept Videos
Guidelines and Strategies for Safe Computer Charting
Maintain Confidentiality and Security:
Guidelines for Nursing Documentation I
Factual:
The following points emphasize the significance of upholding accurate and unbiased documentation in healthcare.
Methods of Documentation VII: EMR
One-Compartment Model: IV Infusion
The one-compartment model for IV infusion uses mathematical equations to describe the rate of change in drug quantity in the body. At steady-state or infusion equilibrium, the drug input...
Errors occurring during blood pressure monitoring
Several factors...
Legal Guidelines for Documentation