Oscillation and Reaction Board Techniques for Estimating Inertial Properties of a Below-knee Prosthesis
Published on: May 8, 2014
Volume Fluctuations in Active and Nonactive Transtibial Prosthetics Users
Nur Afiqah Hamzah1, Nasrul Anuar Abd Razak1, Mohd Sayuti Ab Karim2
1Department of Biomedical Engineering, Faculty of Engineering, University of Malaya, Kuala Lumpur, Malaysia.
The Biosculptor Bioscanner system accurately measures transtibial residual limb volume fluctuations in amputees. This device shows high correlation and agreement with manual measurements, making it a reliable alternative tool.
Area of Science:
- Biomedical Engineering
- Prosthetics and Orthotics
- Rehabilitation Technology
Background:
- Residual limb volume management is crucial for prosthetic fitting and comfort.
- Understanding limb volume fluctuations during activity is essential for optimizing prosthetic use.
- Current measurement methods may lack precision or convenience for daily monitoring.
Purpose of the Study:
- To evaluate the validity and reliability of the Biosculptor Bioscanner system.
- To assess the Bioscanner's ability to capture transtibial residual limb volume changes.
- To compare Bioscanner measurements with traditional manual measurements.
Main Methods:
- Residual limb volume was measured using the Bioscanner and a tape measure.
- Measurements were taken after amputees walked for 5 to 25 minutes over five consecutive days.
- Circumference and volume calculations were compared between the two methods.
Main Results:
- The Bioscanner showed a high correlation (0.97-1) with manual measurements.
- Percentage differences in volume calculation ranged from 2.4% to 9.3%.
- Limb circumference changes were non-uniform, with significant distal increases after initial walking.
Conclusions:
- The Bioscanner system demonstrates comparable accuracy to standard measurement methods.
- The Bioscanner is a potentially interchangeable and reliable tool for managing daily residual limb volume changes.
- This technology may offer an alternative for monitoring amputees' residual limb volume.