Related Experiment Video
Updated: Aug 28, 2025

Characterization of Recombination Effects in a Liquid Ionization Chamber Used for the Dosimetry of a Radiosurgical Accelerator
Published on: May 9, 2014
SUBSTANTIAL EXTERNAL DOSE RATE VARIABILITY OBSERVED IN A COHORT OF LU-177 PATIENTS INDEPENDENT OF BMI AND SEX
Michael Bellamy1, Bae Chu1, Brian Serencsits1
1Department of Medical Physics, Memorial Sloan Kettering, 1275 York Avenue, New York, NY 10065, USA.
Abstract:
External dose rates were measured 1 m away from 230 Lu-177 patients to characterise the variability in normalised dose rates as a function of administered activity, body mass index (BMI) and sex. The largest dose rate observed was 0.07 mSv/h associated with an administered activity of 7.2 GBq. Substantial variability was found in the distribution of the normalised dose rate associated that had an average of 0.0037 mSv/h per GBq and a 95% confidence interval of 0.0024-0.0058 mSv/h per GBq. Based on this study, estimating the patient dose rate based on the Lu-177 gamma exposure factor overestimates the dose rate by a factor of 2. A statistically significant inverse relationship was found between the patient dose rate and patient BMI and an empirically derived equation relating these two quantities was reported. On average, male patient dose rates were 3.5% lower than female dose rates, which may be attributed to the larger average BMI of the male patient group.
More Related Videos
06:20Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
Published on: March 11, 2021
09:49A Whole Body Dosimetry Protocol for Peptide-Receptor Radionuclide Therapy PRRT: 2D Planar Image and Hybrid 2D+3D SPECT/CT Image Methods
Published on: April 24, 2020
Related Concept Videos
Biological Effects of Radiation
Dose-Response Relationship: Overview