Patient Directed Recording of a Bipolar Three-Lead Electrocardiogram using a Smartwatch with ECG Function
Published on: December 11, 2019
Comparison Between QT and Corrected QT Interval Assessment by an Apple Watch With the AccurBeat Platform and by a
Sara Chokshi1, Gulzhan Tologonova2, Rose Calixte3
1Healthcare Innovation and Technology Lab, New York, NY, United States.
Background:
Abnormal prolongation or shortening of the QT interval is associated with increased risk for ventricular arrhythmias and sudden cardiac death. For continuous monitoring, widespread use, and prevention of cardiac events, advanced wearable technologies are emerging as promising surrogates for conventional 12‑lead electrocardiogram (ECG) QT interval assessment. Previous studies have shown a good agreement between QT and corrected QT (QTc) intervals measured on a smartwatch ECG and a 12-lead ECG, but the clinical accuracy of computerized algorithms for QT and QTc interval measurement from smartwatch ECGs is unclear.
Objective:
The prospective observational study compared the smartwatch-recorded QT and QTc assessed using AccurKardia's AccurBeat platform with the conventional 12‑lead ECG annotated manually by a cardiologist.
Methods:
ECGs were collected from healthy participants (without any known cardiovascular disease) aged >22 years. Two consecutive 30-second ECG readings followed by (within 15 minutes) a 10-second standard 12-lead ECG were recorded for each participant. Characteristics of the participants were compared by sex using a 2-sample t test and Wilcoxon rank sum test. Statistical comparisons of heart rate (HR), QT interval, and QTc interval between the platform and the 12-lead ECG, ECG lead I, and ECG lead II were done using the Wilcoxon sign rank test. Linear regression was used to predict QTc and QT intervals from the ECG based on the platform's QTc/QT intervals with adjustment for age, sex, and difference in HR measurement. The Bland-Altman method was used to check agreement between various QT and QTc interval measurements.
Results:
A total of 50 participants (32 female, mean age 46 years, SD 1 year) were included in the study. The result of the regression model using the platform measurements to predict the 12-lead ECG measurements indicated that, in univariate analysis, QT/QTc intervals from the platform significantly predicted QT/QTc intervals from the 12-lead ECG, ECG lead I, and ECG lead II, and this remained significant after adjustment for sex, age, and change in HR. The Bland-Altman plot results found that 96% of the average QTc interval measurements between the platform and QTc intervals from the 12-lead ECG were within the 95% confidence limit of the average difference between the two measurements, with a mean difference of -10.5 (95% limits of agreement -71.43, 50.43). A total of 94% of the average QT interval measurements between the platform and the 12-lead ECG were within the 95% CI of the average difference between the two measurements, with a mean difference of -6.3 (95% limits of agreement -54.54, 41.94).
Conclusions:
QT and QTc intervals obtained by a smartwatch coupled with the platform's assessment were comparable to those from a 12-lead ECG. Accordingly, with further refinements, remote monitoring using this technology holds promise for the identification of QT interval prolongation.
Related Concept Videos
Electrocardiogram
Three major waveforms are present in a typical ECG recording: the P wave, the QRS complex, and...
Electrocardiogram Fundamentals
An electrocardiogram (ECG) is a diagnostic tool for identifying cardiac conditions such as arrhythmias, conduction abnormalities, and myocardial ischemia.
Definition
An electrocardiogram (ECG) visualizes the heart's electrical activity by tracing the electrical movement associated with each heartbeat on a graph or monitor. As the heart beats, an electrical wave passes through it, correlating with the cardiac cycle events.
Parts of an ECG
An ECG utilizes electrodes on the skin...
ECG Interpretation of Rhythms
Components of the Electrocardiogram
The primary components of a normal ECG waveform in Normal sinus rhythm(NSR) include the P wave, PR interval, QRS complex, ST segment, T wave, and occasionally a U wave.
ECG waveforms are divided by vertical and horizontal lines at standard intervals.
The horizontal axis measures time and rate, and the vertical axis measures amplitude or voltage....
Correlation between ECG and Cardiac Cycle
A cardiac action potential originates in the SA node and spreads throughout the atria and the AV node in approximately 0.03 seconds. This results in the P wave in an ECG and triggers atrial contraction. The action potential is then briefly slowed at the AV node, allowing the atria to contract and fill the ventricles with blood before...
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Holter Monitor: 24-Hour Monitoring