Prediction of atrial fibrillation using a home blood pressure monitor with a high-resolution system
Yasunaga Hiyoshi1, Hidenobu Hashimoto2, Takayuki Kabuki2
1Department of Cardiology, Tokyo Metropolitan Ebara Hospital, Ota-ku, Tokyo, Japan yasunaga_hiyoshi@tmhp.jp.
Objective:
The usefulness of screening for atrial fibrillation (AF) using several home blood pressure (BP) monitors has been reported. We evaluated the accuracy of a high-resolution system (HiRS) for AF prediction and its usefulness when installed in home BP monitors.
Methods:
In patients with paroxysmal, persistent or permanent AF, ECG recording and BP measurements were performed simultaneously. The relationship between ECG rhythm diagnosis and pulse irregularity recognition, using a home BP monitor with HiRS, was investigated. The severity of a pulse disturbance during BP measurement was displayed as an irregular pulse rhythm symbol (IPRS) in three instances. The IPRS was not displayed if the pulse was regular, turned on if there was a weak variation in the pulse, and blinked if there was a strong variation in the pulse.
Results:
One hundred and seven patients (44 paroxysmal AF, 63 persistent or permanent AF) were enrolled, and a total of 333 recordings were analysed. The rhythms recorded by each ECG were 73 sinus regular rhythms, 35 extrasystoles, 222 AFs and 3 atrial flutters. Sensitivity and specificity for the prediction of any arrhythmia by the IPRS display of the BP monitor were 95.8% (95% CI 92.6% to 97.6%) and 96.8% (95% CI 92.6% to 100%), respectively. In addition, sensitivity and specificity for the prediction of AF were 100% (95% CI 97.5% to 100%) and 74.8% (95% CI 65.6% to 82.5%), respectively. Sensitivity and specificity for the prediction of AF by the IPRS blinking display were 88.3% (95% CI 83.3% to 92.2%) and 94.6% (95% CI 88.6% to 98.0%%), respectively. IPRS exhibited lighting or blinking during AF occurrence; however, during sinus rhythm, IPRS was not displayed in 72 out of 73 recordings.
Conclusion:
The IPRS device predicted AF with precision and may be particularly useful for predicting an arrhythmia attack in patients with paroxysmal AF.
More Related Videos
09:17High-Resolution Endocardial and Epicardial Optical Mapping in a Sheep Model of Stretch-Induced Atrial Fibrillation
Published on: July 29, 2011
08:10Estimating Bilateral Atrial Function by Cardiovascular Magnetic Resonance Feature Tracking in Patients with Paroxysmal Atrial Fibrillation
Published on: July 20, 2022
Related Concept Videos
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Errors occurring during blood pressure monitoring
Several factors...
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Measurement of Blood Pressure
Assessment of blood pressure in brachial artery(two-step method)