Design and Implementation of a Bespoke Robotic Manipulator for Extra-corporeal Ultrasound
Published on: January 7, 2019
Optimisation and validation of a co-manipulated robot for brachytherapy procedure
Aziza Ben Halima1, Julien Bert1, Jean-François Clément1
1Laboratory of Medical Image Processing (LaTIM), INSERM UMR 1011, University of Western Brittany (UBO), Brest, France.
Background:
Low-dose rate brachytherapy is the referent treatment for early-stage prostate cancer and consists in manually inserting radioactive seeds within the organ to destroy tumorous cells. This treatment is inaccurate leading to side effects. Researchers developed robots to improve this technique. Despite ameliorating accuracy, they cannot be clinically used because of size and acceptability. Therefore, a 6-DOF parallel and co-manipulated robot is proposed to meet these requirements.
Methods:
To fulfil the application requirements, a compact design was modelled. The robot's optimal dimensions were defined by establishing kinematics and implementing genetic algorithm. The robot's relevance was evaluated by measuring workspace and needle placement errors.
Results:
The robot fits into a cube of 300 × 300 × 300 mm3 and provides a free-singularity workspace of 55 × 55 × 150 mm3 with a possible end-effector rotation of 15° and a needle placement error <3 mm.
Conclusion:
The results are promising and prove that our robot fulfils the application requirements and presents a beneficial alternative to the manual procedure.