Quantification of Oculomotor Responses and Accommodation Through Instrumentation and Analysis Toolboxes
Published on: March 3, 2023
Comparison of wavefront aberrations in the object and image spaces using wide-field individual eye models
Yongji Liu1,2, Xiaolan Li1, Lin Zhang2,3
1Tianjin Key Laboratory of Micro-scale Optical Information Science and Technology, Institute of Modern Optics, Nankai University, 38 Tongyan Road, Haihe Education Park, Tianjin, 300350, China.
Abstract:
Wavefront aberrations in the image space are critical for visual perception, though the clinical available instruments usually give the wavefront aberrations in the object space. This study aims to compare the aberrations in the object and image spaces. With the measured wavefront aberrations over the horizontal and vertical ±15° visual fields, the in-going and out-going wide-field individual myopic eye models were constructed to obtain the wavefront aberrations in the object and image spaces of the same eye over ±45° horizontal and vertical visual fields. The average differences in the mean sphere and astigmatism were below 0.25 D between the object and image spaces over the horizontal and vertical ±45° visual fields under 3 mm and 6 mm pupil diameter. The wavefront aberrations in the object space are a proper representation of the aberrations in the image space at least for horizontal visual fields ranging from -35°to +35° and vertical visual fields ranging from -15°to +15°.