Evaluation of Gas Chromatography Columns with Radially Elongated Pillars as Second-Dimension Columns in Comprehensive
Amel Meziani1,2, Sandrien Verloy3,4, Ouassila Ferroukhi2
1Univ Rouen Normandie, FR CNRS 3038, SMS, UR3233, Rouen F-76000, France.
Abstract:
The present study investigated the use of a dedicated gas chromatography (GC) column (L = 70 cm, 75 μm deep, and 6.195 mm wide) with radially elongated pillars (REPs) as the second column in a comprehensive two-dimensional gas chromatography (GC × μGC) system. Three stationary phases [apolar polydimethylsiloxane (PDMS), medium polar room-temperature ionic liquid (RTIL) based on monocationic phosphonium, and polar polyethylene glycol (PEG-1000)] have been coated using the static method at constant pressure or using an original vacuum pressure program (VPP) from 400 to 4 mbar. The best efficiency reached up to N = 62,000 theoretical plates for a film thickness of 47 nm at 100 °C for an iso-octane peak (k = 0.16) at an optimal flow rate of 4.8 mL/min. The use of the VPP improved the efficiency by approximately 15%. Efficiencies up to 28,000 and 47,000 were obtained for PEG-1000 and RTIL, respectively. A temperature-programmed separation of a mixture of 11 volatile compounds on a PDMS-coated chip was obtained in less than 36 s. The PDMS-, PEG-1000-, and RTIL-coated chips were tested as the second column using a microfluidic reverse fill/flush flow modulator in a GC × μGC system. The REP columns were highly compatible with the operating conditions in terms of flow rate and with more than 30,000 plates for the iso-octane peak. Moreover, a commercial solvent called white spirit containing alkanes and aromatic compounds was injected in three sets of columns in normal and reverse modes, demonstrating the great potential of the chip as a second-dimension separation column.
More Related Videos
Related Concept Videos
Gas Chromatography: Types of Columns and Stationary Phases
For an analyte to remain on the column for a sufficient amount of time, it must exhibit some level of compatibility (or...
Gas Chromatography: Introduction
In GC, a sample is vaporized and mixed with an inert carrier gas (the mobile phase), which transports it through a...
Diffusion on Chromatography Columns
Longitudinal diffusion occurs when the solute molecules in the mobile phase diffuse from the more concentrated center of the chromatographic band to the more dilute regions on either side, both towards and against the flow direction. This...
Types Of Column Chromatography
Gel Filtration Chromatography
When the...
Principles Of Column Chromatography
Gas Chromatography–Mass Spectrometry (GC–MS)
A gas chromatograph consists of a long, narrow capillary column with a polysiloxane coating on the inner wall....