Comparison of Performance between the Elecsys 4th and 5th Generation STAT Cardiac Troponin T Immunoassays
Mustafa A Barbhuiya1, Terri Neibauer1, Eric Saylor1
1Department of Pathology and Laboratory Medicine, Milton S. Hershey Medical Center and Pennsylvania State University College of Medicine, Hershey, PA, USA.
Objective:
To evaluate the analytical and clinical performance of the 5th GEN cTnT assay by comparing it with the 4th GEN assay.
Methods:
Imprecision, analytical measurement range (AMR), reference interval, and quantitative comparison were studied. Qualitative comparisons of the two assays for randomly selected patients with the cTnT test orders and patients with elevated N-terminal proB-type natriuretic peptide (NT-proBNP), decreased estimated glomerular filtration rate (eGFR), and increased procalcitonin were performed.
Results:
Within-run and between-run CVs of the 5th GEN assay were 0.6-2.3% and 3.1-5.0% with AMR of 8-10,000 ng/L and reference interval of <19 ng/L. Regression analysis of the two assays showed linear relationship: y (5th GEN)=0.899x (4th GEN)+26, r=0.9993 with a positive bias of the 5th GEN assay in samples with cTnT <150 ng/L and a negative bias in samples with cTnT >500 ng/L compare to the 4th GEN assay. Agreement was 88.1% (95% CI: 80.4 to 93.1%) between the two methods according to the 99th percentile threshold. In patients with elevated NT-proBNP, decreased eGFR, and increased procalcitonin, agreements were 79.4%, 92.7%, and 100%.
Conclusion:
The 5th GEN cTnT assay demonstrates excellent precision and acceptable AMR. Compared to the 4th GEN assay, the 5th GEN assay shows a negative bias for samples with cTnT >500 ng/L but a positive bias for samples with cTnT <150 ng/L and identifies more patients with elevated cTnT.
More Related Videos
07:40Impact of High-intensity Interval Exercise and Moderate-Intensity Continuous Exercise on the Cardiac Troponin T Level at an Early Stage of Training
Published on: October 10, 2019
09:23Using Extraordinary Optical Transmission to Quantify Cardiac Biomarkers in Human Serum
Published on: December 13, 2017
Related Concept Videos
Blood Studies for Cardiovascular System I: Cardiac Biomarkers
The essential diagnostic tools for detecting myocardial necrosis and monitoring individuals suspected of having acute coronary syndrome (ACS) include:
Troponins
Troponins, particularly cardiac troponins I and T, are the most precise and sensitive markers of myocardial injury. They are detectable within 4-6 hours of myocardial injury and remain...
Acute Coronary Syndrome III: Diagnostic Studies
Electrocardiogram
Three major waveforms are present in a typical ECG recording: the P wave, the QRS complex, and...