Speech Recognition Performance Differences Between Precurved and Straight Electrode Arrays From a Single Manufacturer
Rahul K Sharma1, Miriam R Smetak1, Ankita Patro1
1Department of Otolaryngology-Head & Neck Surgery, Vanderbilt University Medical Center.
Summary
Precurved cochlear implant electrode arrays significantly improved speech recognition scores compared to straight arrays, even after accounting for patient age and preoperative hearing. This finding aids in selecting the optimal cochlear implant electrode for better hearing outcomes.
Area of Science:
- Otolaryngology
- Biomedical Engineering
- Audiology
Background:
- Precurved cochlear implant (CI) electrode arrays may offer superior audiometric outcomes compared to straight electrodes.
- Previous studies often lacked control for crucial factors like preoperative hearing and patient age.
- This research addresses these limitations by comparing outcomes in a large cohort while controlling for confounding variables.
Purpose of the Study:
- To compare postoperative speech recognition and audiometric outcomes between precurved and straight cochlear implant electrode arrays.
- To investigate the influence of electrode array type on hearing performance in adult CI recipients.
- To provide data that can guide the selection of electrode arrays for improved patient results.
Main Methods:
- A retrospective cohort study was conducted involving 231 adult CI recipients implanted between 2015 and 2021.
- Patients received either straight (522/622) or precurved (532/632) electrode arrays from a single manufacturer.
- Speech recognition (CNCw, AzBio) and hearing preservation were assessed at 6 and 12 months postactivation, with statistical models controlling for age, race, sex, preoperative hearing, and follow-up time.
Main Results:
- Precurved electrode arrays demonstrated significantly better performance on Consonant-Nucleus-Consonant word (CNCw) scores (p < 0.001) and AzBio scores (p = 0.014) compared to straight arrays, after adjusting for covariates.
- While AzBio scores showed a trend towards significance (p = 0.081) pre-adjustment, the adjusted analysis confirmed a significant benefit for precurved arrays.
- No significant difference in hearing preservation was observed between the two electrode types in the adjusted models.
Conclusions:
- Adult cochlear implant recipients with precurved electrode arrays achieved significantly higher speech recognition scores (CNCw and AzBio) than those with straight arrays.
- These improved outcomes were observed even when controlling for key factors such as age, preoperative hearing levels, and follow-up duration.
- The findings support the use of precurved electrode arrays for potentially enhancing speech understanding in cochlear implant users.