Evaluation of Surface Dose and Commissioning of Compensator-Based Total Body Irradiation
Bharath Pandu1,2, D Khanna1, P Mohandass3
1Department of Applied Physics, Karunya Institute of Technology and Sciences, Coimbatore, Tamil Nadu, India.
Purpose:
The aim of the current study is to commission compensator-based total body irradiation (TBI) and to compare surface dose using percentage depth dose (PDD) while varying the distance between beam spoiler and phantom surface.
Materials And Methods:
TBI commissioning was performed on Elekta Synergy® Platform linear accelerator for bilateral extended source to surface distance treatment technique. The PDD was measured by varying the distance (10 cm, 20 cm, 30 cm, and 40 cm) between the beam spoiler and the phantom surface. Beam profile and half-value layer (HVL) measurement were carried out using the FC65 ion-chamber. Quality assurance (QA) was performed using an in-house rice-flour phantom (RFP). In-vivo diodes (IVD) were placed on the RFP at various regions to measure the delivered dose, and it was compared to the calculated dose.
Results:
An increase in Dmax and surface dose was observed when beam spoiler was moved away from the phantom surface. The flatness and symmetry of the beam profile were calculated. The HVL of Perspex and aluminum is 17 cm and 8 cm, respectively. The calculated dose of each region was compared to the measured dose on the RFP with IVD, and the findings showed that the variation was <4.7% for both Perspex and Aluminum compensators.
Conclusion:
The commissioning of the compensator-based TBI technique was performed and its QA measurements were carried out. The Mayneord factor corrected PDD and measured PDD values were compared. The results are well within the clinical tolerance limit. This study concludes that 10 cm -20 cm is the optimal distance from the beam spoiler to phantom surface to achieve prescribed dose to the skin.
More Related Videos
07:31Characterization of Recombination Effects in a Liquid Ionization Chamber Used for the Dosimetry of a Radiosurgical Accelerator
Published on: May 9, 2014
07:57Positron Emission Tomography-based Dose Painting Radiation Therapy in a Glioblastoma Rat Model using the Small Animal Radiation Research Platform
Published on: March 24, 2022
Related Concept Videos
Biological Effects of Radiation
Radiation: Applications
The average...
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Absorption of Radiation
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET