Related Experiment Video
Updated: Aug 26, 2025

Cardiac Magnetic Resonance for the Evaluation of Suspected Cardiac Thrombus: Conventional and Emerging Techniques
Published on: June 11, 2019
Safety, feasibility, and prognostic value of stress perfusion CMR in patients with MR-conditional pacemaker
Théo Pezel1,2,3, Jérôme Lacotte4, Jérôme Horvilleur4
1Université de Paris Cité, Department of Cardiology, Hôpital Lariboisière - APHP, Inserm UMRS 942, 75010 Paris, France.
Aims:
To assess the safety, feasibility, and prognostic value of stress cardiovascular magnetic resonance (CMR) in patients with pacemaker (PM).
Methods And Results:
Between 2008 and 2021, we conducted a bi-centre longitudinal study with all consecutive patients with MR-conditional PM referred for vasodilator stress CMR at 1.5 T in the Institut Cardiovasculaire Paris Sud and Lariboisiere University Hospital. They were followed for the occurrence of major adverse cardiovascular events (MACE) defined as cardiac death or non-fatal myocardial infarction. Cox regression analyses were performed to determine the prognostic value of CMR parameters. The quality of CMR was rated by two observers blinded to clinical details. Of 304 patients who completed the CMR protocol, 273 patients (70% male, mean age 71 ± 9 years) completed the follow-up (median [interquartile range], 7.1 [5.4-7.5] years). Among those, 32 experienced a MACE (11.7%). Stress CMR was well tolerated with no significant change in lead thresholds or pacing parameters. Overall, the image quality was rated good or excellent in 84.9% of segments. Ischaemia and late gadolinium enhancement (LGE) were significantly associated with the occurrence of MACE (hazard ratio, HR: 11.71 [95% CI: 4.60-28.2]; and HR: 5.62 [95% CI: 2.02-16.21], both P < 0.001). After adjustment for traditional risk factors, ischaemia and LGE were independent predictors of MACE (HR: 5.08 [95% CI: 2.58-14.0]; and HR: 2.28 [95% CI: 2.05-3.76]; both P < 0.001).
Conclusion:
Stress CMR is safe, feasible and has a good discriminative prognostic value in consecutive patients with PM.
More Related Videos
05:07Author Spotlight: Improved Localization and Monitoring of Coronary Flow Reserve Using Modified PLAX View in Mice
Published on: August 25, 2023
11:13Quantification of Mouse Heart Left Ventricular Function, Myocardial Strain, and Hemodynamic Forces by Cardiovascular Magnetic Resonance Imaging
Published on: May 24, 2021
Related Concept Videos
Imaging Studies for Cardiovascular System IV: CMRI
Mitral Regurgitation II: Clinical Features and Diagnostic Tests
Imaging Studies for Cardiovascular System V: CT
Mitral Stenosis II: Clinical features and Diagnostic Tests