Automated bacteriuria screening using the Berthold LB 950 luminescence analyser
European Journal of Clinical Microbiology
|June 1, 1987
Summary
This study shows an automated bioluminescence test accurately screens for bacteriuria (bacterial infection in urine) in 15 minutes. The test is highly sensitive and specific, especially with uncontaminated samples.
Area of Science:
- Clinical Microbiology
- Biochemistry
Background:
- Bacteriuria diagnosis traditionally relies on urine culture, which is time-consuming.
- Rapid and automated methods are needed for efficient screening of bacterial infections in urine.
Purpose of the Study:
- To evaluate the performance of an automated luminescence analyzer for estimating bacterial adenosine triphosphate (ATP) in urine.
- To assess the utility of this automated method as a rapid screening test for bacteriuria.
Main Methods:
- The Berthold LB950 Automatic Luminescence Analyser was used to measure bacterial ATP in 1040 urine samples.
- Bioluminescence results were compared against two reference culture methods (pour plate and standard loop).
- Frequency distributions of bacterial counts and ATP levels were analyzed.
Main Results:
- The automated test provided results within 15 minutes, detecting bacteriuria at the 10^5 CFU/ml level.
- Specificities ranged from 79% to 98% (when contaminated urines were excluded).
- Sensitivities were 84% (pour plate) and 91% (standard loop), improving to 98% with uncontaminated samples.
Conclusions:
- The automated bioluminescence assay is a rapid and effective screening tool for bacteriuria.
- The method demonstrates good sensitivity and specificity, particularly for high-quality urine specimens.
- While the automated test is expensive, it offers significant time advantages over traditional culture methods.