Precise Limb Tourniquet Arterial Occlusion Pressure Determination using Real-Time Ultrasonography and a
Jeffrey N Wood1,2, Benjamin S Krippendorf2, Craig A Blakeney1,2
1Division of Prehospital Care, Mayo Clinic, Rochester, Minnesota, USA.
Background:
Hemorrhage control prior to shock onset is increasingly recognized as a time-critical intervention. Although tourniquets (TQs) have been demonstrated to save lives, less is known about the physiologic parameters underlying successful TQ application beyond palpation of distal pulses. The current study directly visualized distal arterial occlusion via ultrasonography and measured associated pressure and contact force.
Methods:
Fifteen tactical officers participated as live models for the study. Arterial occlusion was performed using a standard adult blood pressure (BP) cuff and a Combat Application Tourniquet Generation 7 (CAT7) TQ, applied sequentially to the left mid-bicep. Arterial flow cessation was determined by radial artery palpation and brachial artery pulsed wave doppler ultrasound (US) evaluation. Steady state maximal generated force was measured using a thin-film force sensor.
Results:
The mean (95% CI) systolic blood pressure (SBP) required to occlude palpable distal pulse was 112.9mmHg (109-117); contact force was 23.8N [Newton] (22.0-25.6). Arterial flow was visible via US in 100% of subjects despite lack of palpable pulse. The mean (95% CI) SBP and contact force to eliminate US flow were 132mmHg (127-137) and 27.7N (25.1-30.3). The mean (95% CI) number of windlass turns to eliminate a palpable pulse was 1.3 (1.0-1.6) while 1.6 (1.2-1.9) turns were required to eliminate US flow.
Conclusions:
Loss of distal radial pulse does not indicate lack of arterial flow distal to upper extremity TQ. On average, an additional one-quarter windlass turn was required to eliminate distal flow. Blood pressure and force measurements derived in this study may provide data to guide future TQ designs and inexpensive, physiologically accurate TQ training models.
Related Concept Videos
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Sites for measruring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessment of blood pressure in brachial artery(two-step method)