CONTROL FACTORS FOR SITE ERRORS MANAGEMENT OF RADIOTHERAPY DELIVERY
Hendrik Hendrik1, Massila Kamalrudin2, Mohamad Razali2
1DR.MOEWARDI GENERAL HOSPITAL, SURAKARTA, INDONESIA; UNIVERSITAS SEBELAS MARET, SURAKARTA, CENTRAL OF JAVA, INDONESIA.
Objective:
The aim: This study aimed to define the factors related to irradiation field equality and target accuracy which will further influence the irradiation result.
Patients And Methods:
Materials and methods: This is a prospective-qualitative study, conducted by observation of image data verification from cervical cancer patients in the Department of Radiotherapy, dr. Moewardi Hospital, Surakarta who had undergone several times a week irradiation utilizing Telecobalt60 device and by conducting an in-depth interview to ten Indonesian radiotherapy experts, in October 2018. The data was further analyzed using correlation - regression test.
Results:
Results: From 30 verification image data of the irradiated patients, we conclude that the scheme, body size, and patient positioning factors have all revealed statistically significant correlations to the irradiation field equality. On the other hand, factors such as patient and tele-therapy device set-ups, tele-therapy device calibration, human resources quality, and tele-therapy device malfunctions have all revealed statistically significant correlations to the irradiation target accuracy. These facts were further strengthen by the supporting statements from 10 Indonesian radiotherapy experts.
Conclusion:
Conclusions: The impact factors of field equality and accuracy of the irradiation target could serve as an important control factors which is substantially required to manage and minimize site errors of the radiotherapy delivery.
Related Concept Videos
Types of Errors: Detection and Minimization
Absolute error in a measurement is the numerical difference from the true or central value. Relative error is the ratio between absolute error and the true or central value, expressed as a percentage.
Errors can be classified by source, magnitude, and sign. There are three types of errors: systematic, random, and gross.
Systematic or...
Common Leveling Mistakes and Errors
Errors in Taping
Radiation Pressure: Problem Solving
The average value of the rate of momentum transfer divided by the absorbing area represents the average force...
Errors occurring during blood pressure monitoring
Several factors...
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...