A novel approach for Venezia ovoid commissioning with a comprehensive analysis of source positions in high-dose-rate
Xing Li1, Jessica Huang1, Robabeh Rahimi2
1University of Utah, Department of Radiation Oncology, Salt Lake City, UT.
Purpose:
The lunar design of a Venezia ovoid makes commissioning of the applicator very challenging with traditional autoradiography. In this study, we propose a novel solution to ovoid commissioning and a quality assurance (QA) workflow to effectively assess the entire source path.
Methods And Materials:
A two-step commissioning process, using electron radiation and radiochromic films, was developed to verify the most distal source position. The ovoid was first attached to a film and was irradiated with a 12 MeV linac beam. This process was repeated on a separate, unexposed film, followed by irradiating it with a HDR source at the most distal position. Two lengths, including the ovoid thickness and the distance between the irradiated spot and the ovoid's outer surface, were obtained from the films' intensity maps. The offset value was calculated from the subtraction of the two measured lengths. Besides acquiring the offset, a source positional simulator (SPS) and a series of planar x-rays from two orthogonal orientations were used to characterize source movement within the ovoid.
Results:
Compared to x-ray-based autoradiography, the electron exposure significantly improved the ovoid's visibility on film. Our approach did not use surrogate, which further improved measurement outcomes by decreasing inherent uncertainties. The SPS results suggested the source movement was complex within the cervicovaginal area, but it was predictable with the proposed QA workflow.
Conclusion:
We introduced a novel, surrogate-free method to commission the Venezia ovoid, which facilitated a manual applicator reconstruction. Additionally, we recommended QA multiple source positions to safely use the ovoid in clinical settings.
More Related Videos
06:20Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
Published on: March 11, 2021
07:57Positron Emission Tomography-based Dose Painting Radiation Therapy in a Glioblastoma Rat Model using the Small Animal Radiation Research Platform
Published on: March 24, 2022
Related Concept Videos
One-Compartment Open Model for IV Bolus Administration: General Considerations
The drug's presence in the body is defined by an equation representing the difference between the rates of drug entry and exit. Key parameters—elimination rate constant,...
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
One-Compartment Open Model for IV Bolus Administration: Estimation of Clearance
In the one-compartment open model for intravenous (IV) bolus administration, clearance is estimated by dividing the elimination rate by the plasma drug concentration. This equation leverages the elimination rate constant and the apparent...
Radiation: Applications
The average...
Radiological Investigation II: MRI and Ventilation Perfusion Scan
Magnetic Resonance Imaging (MRI) and Ventilation Perfusion Scans are two radiological investigations that offer detailed diagnostic images of the body, particularly lung structures.
MRI
MRI uses magnetic fields and radiofrequency signals to distinguish between normal and abnormal tissues. This technology provides a more detailed diagnostic image than CT scans, enabling it to characterize pulmonary nodules, stage bronchogenic carcinoma, and evaluate inflammatory activity in...
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...