Comparative retrospective study on the validity of point-of-care testing device for massive obstetrical hemorrhage:
Eishin Nakamura1, Shigetaka Matsunaga2, Akihiko Kikuchi1
1Center for Maternal, Fetal and Neonatal Medicine, Saitama Medical Center, Saitama Medical University, Saitama, Japan (Dr Nakamura and Dr Kikuchi).
Background:
Early recognition of hypofibrinogenemia and prompt initiation of transfusion therapy in patients with massive obstetrical hemorrhage can improve prognosis. There are reports on the usefulness of point-of-care testing, which provides quicker test results compared with fibrinogen measurements using the conventional Clauss method.
Objective:
This study aimed to compare and investigate the diagnostic accuracy of dry hematology and thromboelastography in point-of-care testing for the diagnosis of hypofibrinogenemia.
Study Design:
A single-center, retrospective study of 126 massive obstetrical hemorrhage cases with point-of-care testing before treatment was initiated. The correlation of fibrinogen values with the Clauss method and the diagnostic accuracy for hypofibrinogenemia were compared between dry hematology and thromboelastography.
Results:
Fibrinogen value in dry hematology showed a strong positive correlation with values measured by the Clauss method, and the diagnostic accuracy for hypofibrinogenemia was high, but there were many residuals above 100 mg/dL, and the distribution of these residuals was not uniform. Although thromboelastography cannot be used to directly measure fibrinogen values, maximum amplitude citrated functional fibrinogen, amplitude-10 citrated rapid thromboelastography, and amplitude-10 citrated functional fibrinogen showed a strong positive correlation with fibrinogen values using the Clauss method, and no significant difference in correlation or diagnostic accuracy was observed relative to dry hematology.
Conclusion:
Dry hematology and thromboelastography were equally accurate in diagnosing hypofibrinogenemia, with results correlating well with fibrinogen values measured by the Clauss method.