Core Lab Adjudication of the ACURATE neo2 Hemodynamic Performance Using Computed-Tomography-Corrected Left
Ahmed Elkoumy1,2, Andreas Rück3, Won-Keun Kim4
1Discipline of Cardiology, Saolta Group, Galway University Hospital, Health Service Executive and CORRIB Core Lab, University of Galway, H91 V4AY Galway, Ireland.
Accurate hemodynamic assessment of prosthetic heart valves requires 3D-MDCT to correct left ventricular outflow tract (LVOT) measurements, improving prosthesis-patient mismatch diagnosis compared to 2D-TTE alone.
Area of Science:
- Cardiology
- Medical Imaging
- Biomedical Engineering
Background:
- Conventional 2D transthoracic Echocardiography-Doppler (2D-TTE) has limitations in assessing prosthetic heart valve hemodynamics, particularly in measuring the left ventricular outflow tract (LVOT) area, which assumes a circular shape.
- Accurate hemodynamic assessment is crucial for evaluating outcomes after transcatheter aortic valve implantation (TAVI).
- Prosthesis-patient mismatch (PPM) can affect long-term outcomes and is influenced by accurate LVOT area measurement.
Purpose of the Study:
- To compare the accuracy of 2D-TTE versus 3D-MDCT for measuring LVOT area and its impact on hemodynamic parameters and PPM incidence in patients receiving the ACURATE neo2 TAVI device.
- To evaluate the effectiveness of a corrected continuity equation using 3D-LVOT area from MDCT combined with TTE Doppler parameters for improved hemodynamic assessment.
Main Methods:
- A study involving 258 patients with severe aortic stenosis treated with the ACURATE neo2 valve.
- LVOT area and Doppler-derived parameters (EOA, SV) were calculated using post-TAVI 2D-TTE.
- 3D-LVOT area was obtained from pre-procedural MDCT scans to calculate corrected Doppler parameters and compare PPM incidence between 2D-TTE and 3D-MDCT methods.
Main Results:
- 2D-TTE significantly underestimated LVOT area (350.4 mm² vs. 405.22 mm²) and consequently EOA and SV compared to 3D-MDCT.
- PPM incidence was higher with 2D-TTE-based measurements (8.52%) than with 3D-MDCT-corrected measurements (2.32%).
- Significant differences in EOA among ACURATE neo2 valve sizes were observed with MDCT but not with 2D-TTE.
Conclusions:
- Combining 3D-LVOT area from MDCT with TTE Doppler parameters offers a more accurate hemodynamic assessment and PPM diagnosis in TAVI patients.
- The ACURATE neo2 THV demonstrates a large EOA and low PPM incidence when assessed using 3D-corrected LVOT area.
- Further long-term follow-up studies are needed to confirm these findings.
More Related Videos
11:09High-Resolution Cardiac Positron Emission Tomography/Computed Tomography for Small Animals
Published on: December 16, 2022
12:45Benefits of Cardiac Resynchronization Therapy in an Asynchronous Heart Failure Model Induced by Left Bundle Branch Ablation and Rapid Pacing
Published on: December 11, 2017
Related Concept Videos
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT
Imaging Studies for Cardiovascular System V: CT
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Imaging Studies for Cardiovascular System I:Echocardiography
Indications: Echocardiography is utilized to diagnose heart failure, valve disorders, and myocardial infarction. It also assesses cardiac structures' size, shape, and motion,...