Related Experiment Video
Updated: Aug 23, 2025

06:26
Novel and Innovative Hybrid Technique for Type A Aortic Dissection
Published on: March 28, 2025
448
A modified candy-plug technique to occlude false lumen in aortic dissection.
Erik Palm1, Antti Valtola2, Hannu Manninen3,4
1Department of Radiology, Kuopio University Hospital, Puijonlaaksontie 2, 70210, Kuopio, Finland. erik.palm@kuh.fi.
CVIR Endovascular
|October 29, 2022
Summary
A novel candy-plug (CP) device effectively occludes false lumen (FL) backflow in aortic dissection. This modified technique successfully reduced aortic diameter and stopped retrograde FL filling in a patient case.
Area of Science:
- Cardiovascular Surgery
- Endovascular Techniques
- Aortic Disease Management
Background:
- Type A aortic dissection can lead to persistent false lumen (FL) filling and aortic dilation.
- Endovascular Aortic Repair (TEVAR) may not always resolve retrograde FL flow from distal tears.
- Management of residual FL backflow after initial aortic repair presents a clinical challenge.
Purpose of the Study:
- To introduce a modified, custom-created candy-plug (CP) device for endovascular occlusion of false lumen (FL).
- To demonstrate the utility of this novel CP device in a patient with persistent FL backflow after TEVAR.
- To present a technique for managing retrograde FL filling in complex aortic dissection.
Main Methods:
- Modification of a commercially available aortic extender endoprosthesis into a candy-plug (CP) device.
- Endovascular positioning of the custom CP device into the false lumen (FL) of the distal descending thoracic aorta.
- Procedure performed under local anesthesia in a patient with a history of type A aortic dissection and TEVAR.
Main Results:
- Successful occlusion of retrograde false lumen (FL) filling was achieved.
- Aortic diameter decreased from 69 mm to 66 mm at 1-month follow-up.
- False lumen (FL) diameter reduced from 37 mm to 34 mm, while true lumen (TL) remained stable at 32 mm.
Conclusions:
- The modified candy-plug (CP) technique is an effective method for occluding retrograde false lumen (FL) filling in aortic dissection.
- This approach offers a viable solution for managing persistent FL backflow after initial aortic interventions.
- The described technique provides a novel option for improving outcomes in complex aortic dissection cases.