Angiographic findings and treatment of coronary sinus spasm during left ventricular lead placement: A case report
Tomo Komaki1, Yuuka Shibata1, Kaori Mine1
1Department of Cardiology, Fukuoka University School of Medicine, Fukuoka, Japan.
Insights
Coronary sinus spasm is a rare complication during left ventricular lead placement for cardiac resynchronization therapy. Prompt recognition and treatment with nitrates can resolve the spasm, enabling successful lead implantation.
Area of Science:
- Cardiology
- Interventional Electrophysiology
Background:
- Cardiac resynchronization therapy (CRT) with an implantable defibrillator is crucial for patients with heart failure and conduction abnormalities.
- Left ventricular (LV) lead placement via the coronary sinus (CS) is a key component of CRT.
Observation:
- During LV lead placement, a 71-year-old man experienced significant resistance during deep CS cannulation.
- CS venography revealed a complete occlusion, which spontaneously resolved after several minutes, indicating venospasm.
Findings:
- Coronary sinus spasm presented as an abrupt occlusion on venography.
- Intravenous isosorbide dinitrate administration and a brief waiting period without catheter manipulation successfully resolved the spasm.
- Successful LV lead placement was achieved after spasm resolution.
Implications:
- Coronary sinus spasm, though rare, is a potential complication during LV lead placement.
- Awareness of CS spasm's angiographic findings and effective management strategies is crucial for electrophysiologists.
- Timely recognition and treatment can prevent procedural delays and ensure successful CRT device implantation.
Abstract:
A 71-year-old man with ischemic cardiomyopathy, a left ventricular (LV) ejection fraction of 23 %, left bundle branch block with a QRS duration of 160 milliseconds, and nonsustained ventricular tachycardia was admitted for cardiac resynchronization therapy combined with an implantable defibrillator. During LV lead placement, the guiding sheath encountered strong resistance during deep coronary sinus (CS) cannulation. CS venography showed a complete occlusion, and we diagnosed venospasm because the occlusion self-resolved after several minutes. After administering intravenous isosorbide dinitrate and waiting several minutes without manipulating the catheters, we could successfully place the LV lead in the target branch. Although CS spasm is considered a rare complication of LV lead placement, in some cases catheter manipulation can trigger it. Therefore, clinicians should recognize its possibility and be aware of the associated angiographic findings and treatment.
Learning Objective:
In some cases, coronary sinus (CS) spasm can be triggered during left ventricular lead placement. It presents as occlusion with abrupt tapering on CS venography. After administering intravenous nitrates and waiting several minutes without manipulating the catheters, the spasm resolves and the catheter can be advanced.
Related Concept Videos
Cardiac Catheterization III: Left Heart Catheterization
Acute Coronary Syndrome III: Diagnostic Studies
Cardiac Catheterization I: Pre-Procedure Overview
Cardiac Catheterization II: Right Heart Catheterization
Mitral Stenosis II: Clinical features and Diagnostic Tests
Cardiac Catheterization IV: Nursing Management