Related Experiment Video
Updated: Aug 23, 2025

Cooling or Warming the Esophagus to Reduce Esophageal Injury During Left Atrial Ablation in the Treatment of Atrial Fibrillation
Published on: March 15, 2020
Esophageal temperature management during cryoballoon ablation for atrial fibrillation
Joshua Sink1, Kiran Nimmagadda2, Manyun Zhao3
1Department of Internal Medicine, Northwestern University, Chicago, Illinois, USA.
Introduction:
Esophageal thermal injury (ETI) is a well-recognized complication of atrial fibrillation (AF) ablation. Previous studies have demonstrated that direct esophageal cooling reduces ETI during radiofrequency AF ablation. The purpose of this study was to evaluate the use of an esophageal warming device to prevent ETI during cryoballoon ablation (CBA) for AF.
Methods:
This prospective, double-blinded study enrolled 42 patients with symptomatic AF undergoing CBA. Patients were randomized to the treatment group with esophageal warming (42°C) using recirculated water through a multilumen, silicone tube inserted into the esophagus (EnsoETM®; Attune Medical) (WRM) or the control group with a luminal single-electrode esophageal temperature monitoring probe (LET). Patients underwent upper endoscopy esophagogastroduodenoscopy (EGD) the following day. ETI was classified into four grades.
Results:
Baseline patient characteristics were similar between groups. Procedural characteristics including number of freezes, total freeze time, early freeze terminations, coldest balloon temperature, procedure duration, posterior wall ablation, and proton pump inhibitor and transesophageal echocardiogram use before procedure were not different between groups. The EGD was completed in 40/42 patients. There was significantly more ETI in the WRM group compared to the LET group (n = 8 [38%] vs. n = 1 [5%], p = 0.02). All ETI lesions were grade 1 (erythema) or 2 (superficial ulceration). Total freeze time in the left inferior pulmonary vein was predictive of ETI (360 vs. 300 s, p = 0.03).
Conclusion:
Use of a luminal heat exchange tube for esophageal warming during CBA for AF was paradoxically associated with a higher risk of ETI.
More Related Videos
04:58Reduced Procedure Time and Variability with Active Esophageal Cooling During Radiofrequency Ablation for Atrial Fibrillation
Published on: August 25, 2022
05:43Author Spotlight: Improving Lesion Contiguity in Pulmonary Vein Isolation via Proactive Esophageal Cooling
Published on: April 19, 2024
Related Concept Videos
Temperature Measurement Sites
Oral: When assessing oral temperature, the thermometer tip should be placed under the tongue in the posterior sublingual pocket. It offers accurate readings and can be...
Methods of reducing fever
Pharmacological Methods of Reducing Fever:
Cardiopulmonary Resuscitation IV: Pharmacological Management
Assessing Body Temperature - Temporal Artery
Step 1: Perform hand hygiene and don a fresh pair of gloves to prevent cross-infection and ensure patient safety.
Step 2: Explain the procedure to the patient to establish trust. Clear communication establishes trust with the patient, ensures they understand what to expect, promotes cooperation, and enhances comfort during the procedure.
Step 3: Assess the patient's...
Equipments Used to Measure Body Temperature
Glass-bulb Thermometer:
Glass-bulb thermometers are hollow glass tubes with a bulb tip containing liquid such as ethanol or mercury. Historically, glass bulb mercury thermometers were the standard device to measure body temperature. Today, mercury thermometers are prohibited in many countries due to the hazardous effects of mercury and the risk of exposure if the glass bulb breaks. In general,...