An objective comparison of two pulse oximetry sensors with different adhesive systems on healthy human volunteers
Gary Grove1, Timothy Houser2, Jacob Dove3
1cyberDERM, Inc., Broomall, Pennsylvania, USA.
Background:
Medical Adhesive Related Skin Injuries can arise from topically applied medical devices, especially in those with fragile skin, including the elderly and premature infants. The purpose of this study was to compare gentleness and reapplication of two pulse oximetry sensors (OxySoftN and MaxN, Medtronic, Boulder, CO).
Materials And Methods:
Eighteen healthy subjects aged 65 years and older were enrolled in the gentleness trial, and 20 healthy subjects (18-69 years) were enrolled in the reapplication trial. For the gentleness trial, trans-epidermal water loss (TEWL) measurements were made at five sites on each forearm at three time points (baseline [T0], 4-h postinitial wear [T1], 4-h postsecond wear [T2]). Total amount of protein adhered to each device was also determined. For the reapplication trial, a series of 180° peel tests were performed to observe the forces required to detach the sensor from the skin.
Results:
TEWL rates in the tail region were significantly greater with MaxN compared to OxySoftN at T1 (p < 0.05). Both were significantly greater than control (p < 0.05). Further, protein analysis revealed that the amount of protein removed was significantly less with OxySoftN compared to MaxN (p < < 0.0001). Differences in loss of adhesion of the tail region between the two sensors were demonstrated, with OxySoftN depreciating at a much slower rate compared with MaxN.
Conclusion:
The OxySoftN sensor appears to be gentle, even on fragile skin, based on reduced strain on the skin during removal. Further, it demonstrated the ability to withstand several reapplications without functional loss in adhesion.
Related Concept Videos
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is...
Special considerations while measuring pulse
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Temperature Measurement Sites
Oral: When assessing oral temperature, the thermometer tip should be placed under the tongue in the posterior sublingual pocket. It offers accurate readings and can be...