Purpose:
Development of a novel quadrature inductively driven transceive wireless coil for breast MRI at 1.5 T.
Methods:
A quadrature wireless coil (HHMM-coil) design has been developed as a combination of two linearly polarized coils: a pair of 'metasolenoid' coils (MM-coil) and a pair of Helmholtz-type coils (HH-coil). The MM-coil consisted of an array of split-loop resonators. The HH-coil design included two electrically connected flat spirals. All the wireless coils were coupled to a whole-body birdcage coil. The HHMM-coil was studied and compared to the linear coils in terms of transmit and SAR efficiencies via numerical simulations. A prototype of HHMM-coil was built and tested on a 1.5 T scanner in a phantom and healthy volunteer. We also proposed an extended design of the HHMM-coil and compared its performance to a dedicated breast array.
Results:
Numerical simulations of the HHMM-coil with a female voxel model have shown more than a 2.5-fold increase in transmit efficiency and a 1.7-fold enhancement of SAR efficiency compared to the linearly polarized coils. Phantom and in vivo imaging showed good agreement with the numerical simulations. Moreover, the HHMM-coil provided good image quality, visualizing all areas of interest similar to a multichannel breast array with a 32% reduction in signal-to-noise ratio.
Conclusion:
The proposed quadrature HHMM-coil allows the -field to be significantly better focused in the region-of-interest compared to the linearly polarized coils. Thus, the HHMM-coil provides high-quality breast imaging on a 1.5 T scanner using a whole-body birdcage coil for transmit and receive.
Related Concept Videos
Magnetic Resonance Imaging
Magnetic Field Due to Two Straight Wires
Magnetic Field Due To A Thin Straight Wire
Imaging Studies IV: Magnetic Resonance Imaging
Radiological Investigation II: MRI and Ventilation Perfusion Scan
Magnetic Resonance Imaging (MRI) and Ventilation Perfusion Scans are two radiological investigations that offer detailed diagnostic images of the body, particularly lung structures.
MRI
MRI uses magnetic fields and radiofrequency signals to distinguish between normal and abnormal tissues. This technology provides a more detailed diagnostic image than CT scans, enabling it to characterize pulmonary nodules, stage bronchogenic carcinoma, and evaluate inflammatory activity in...
Imaging Studies for Cardiovascular System IV: CMRI