A New Single Chamber Implantable Defibrillator with Atrial Sensing: A Practical Demonstration of Sensing and Ease of Implantation
Published on: February 28, 2012
Perioperative Safety and Early Patient and Device Outcomes Among Subcutaneous Versus Transvenous Implantable
Jeff S Healey1, Andrew D Krahn2, Jamil Bashir2
1Population Health Research Institute, Hamilton, Ontario, Canada (J.S.H., W.F.M., D.P.L., S.F.L., G.W., A.D., S.J.C.).
Background:
Implantable cardioverter defibrillators (ICDs) improve survival in patients at risk for cardiac arrest, but are associated with intravascular lead-related complications. The subcutaneous ICD (S-ICD), with no intravascular components, was developed to minimize lead-related complications.
Objective:
To assess key ICD performance measures related to delivery of ICD therapy, including inappropriate ICD shocks (delivered in absence of life-threatening arrhythmia) and failed ICD shocks (which did not terminate ventricular arrhythmia).
Design:
Randomized, multicenter trial. (ClinicalTrials.gov: NCT02881255).
Setting:
The ATLAS trial.
Patients:
544 eligible patients (141 female) with a primary or secondary prevention indication for an ICD who were younger than age 60 years, had a cardiogenetic phenotype, or had prespecified risk factors for lead complications were electrocardiographically screened and 503 randomly assigned to S-ICD (251 patients) or transvenous ICD (TV-ICD) (252 patients). Mean follow-up was 2.5 years (SD, 1.1). Mean age was 49.0 years (SD, 11.5).
Measurements:
The primary outcome was perioperative major lead-related complications.
Results:
There was a statistically significant reduction in perioperative, lead-related complications, which occurred in 1 patient (0.4%) with an S-ICD and in 12 patients (4.8%) with TV-ICD (-4.4%; 95% CI, -6.9 to -1.9; P = 0.001). There was a trend for more inappropriate shocks with the S-ICD (hazard ratio [HR], 2.37; 95% CI, 0.98 to 5.77), but no increase in failed appropriate ICD shocks (HR, 0.61 (0.15 to 2.57). Patients in the S-ICD group had more ICD site pain, measured on a 10-point numeric rating scale, on the day of implant (4.2 ± 2.8 vs. 2.9 ± 2.2; P < 0.001) and 1 month later (1.3 ± 1.8 vs. 0.9 ± 1.5; P = 0.035).
Limitation:
At present, the ATLAS trial is underpowered to detect differences in clinical shock outcomes; however, extended follow-up is ongoing.
Conclusion:
The S-ICD reduces perioperative, lead-related complications without significantly compromising the effectiveness of ICD shocks, but with more early postoperative pain and a trend for more inappropriate shocks.
Primary Funding Source:
Boston Scientific.
More Related Videos
06:10Utilizing Percutaneous Ventricular Assist Devices in Acute Myocardial Infarction Complicated by Cardiogenic Shock
Published on: June 12, 2021
08:00Reduction of Iatrogenic Atrial Septal Defects with an Anterior and Inferior Transseptal Puncture Site when Operating the Cryoballoon Ablation Catheter
Published on: June 15, 2015
Related Concept Videos
Cardiopulmonary Resuscitation III: AED Use
Cardiac Catheterization IV: Nursing Management