Investigation of arsenic contamination in groundwater using hydride generation atomic absorption spectrometry
Dinh Vu Le1, Pham Thi Kim Giang2, Van Trong Nguyen3
1Faculty of Chemical Engineering, Industrial University of Ho Chi Minh City, 12 Nguyen Van Bao St., Go Vap, Ho Chi Minh City, 70000, Vietnam. ledinhvu@iuh.edu.vn.
Abstract:
The validated hydride generation atomic absorption spectrometry (HG-AAS) method has been used to investigate total arsenic in groundwater. Under optimal experimental conditions, the concentration of arsenic in groundwater can be analysed in the range of 0.5 to 50 µg/L, with a method detection limit of 0.15 µg/L. Its recovery in the field is from 96.3 to 99.8%, with high repeatability. The method was used to observe the total arsenic pollution in groundwater collected in Phu Tho Province, Vietnam. A total of 364 groundwater samples were analysed. The results showed that arsenic pollution was significant, with 15.93% of the samples higher than the maximum permissible level of arsenic. About 20.69% of the contaminated samples had a total arsenic ten times higher (100 µg/L) than the maximum permissible level of arsenic. The pollution source was also considered by comparing the arsenic level in the groundwater with arsenic in the surface water in the same areas. Thus, the use of the high-accuracy and sensitive method, HG-AAS, supplies valuable data on groundwater pollution for water resources management and environmental protection.
Related Concept Videos
Atomic Absorption Spectroscopy: Lab
Solutions containing organic solvents, such as low-molecular-mass alcohols, esters, or ketones, enhance absorbances by increasing...
Atomic Absorption Spectroscopy: Overview
When irradiated by EMR of a particular wavelength, these...
Atomic Absorption Spectroscopy: Interference
Spectral interference occurs when signals from other elements or molecules overlap with the analyte signal, falsely elevating or masking the analyte's absorbance. This interference can be corrected using Zeeman,...
Atomic Absorption Spectroscopy: Atomization Methods
Atomic Absorption Spectroscopy: Instrumentation
The atomizer used in AAS can be either a flame atomizer or an...
Atomic Fluorescence Spectroscopy