Design and Functionality of a Multilumen Thoracic Access Port for Pericardial Access Under Direct Visualization
Jacqueline M Contento1, Paige N Mass1, Rohan N Kumthekar2
1Sheikh Zayed Institute for Pediatric Surgical Innovation, Children's National Hospital, 111 Michigan Avenue NW, Washington, DC 20010.
Journal of Medical Devices
|November 10, 2022
Summary
PeriPath offers a minimally invasive method for cardiac lead implantation, avoiding open-chest surgery. This novel pericardial access tool enhances safety and efficiency for pacing and defibrillation procedures.
Area of Science:
- Cardiovascular Medicine
- Biomedical Engineering
- Surgical Innovation
Background:
- Transvenous cardiac lead implantation is often hindered by vascular anomalies, necessitating invasive open-chest procedures.
- Minimally invasive techniques promise reduced patient trauma, pain, and recovery times, while eliminating fluoroscopy use.
Purpose of the Study:
- To introduce PeriPath, a novel, single-use, low-cost pericardial access tool designed for minimally invasive cardiac lead implantation.
- To evaluate the safety, efficacy, and feasibility of the PeriPath device and procedure.
Main Methods:
- The PeriPath device utilizes a modified Seldinger technique for safe pericardial access and lead placement.
- Mechanical design was validated using finite element analysis (FEA) and tensile strength testing.
- Procedure compatibility with existing cardiac lead implantation tools was assessed.
Main Results:
- FEA confirmed the device's structural integrity under clinical conditions.
- Mechanical testing demonstrated sufficient tensile strength and low fracture risk.
- Preliminary animal and physician studies indicated effective function for minimally invasive lead implantation.
Conclusions:
- PeriPath provides a safe and effective solution for minimally invasive cardiac lead implantation, potentially reducing the need for open-chest surgery.
- The device's design, cost-effectiveness, and compatibility suggest widespread adoption by electrophysiologists and interventional cardiologists.